Service
AIR CELL REPLACEMENT
TOOLS REQUIRED:
●
Soap water and a container
PROCEDURE:
1.
Infalte the air cells. Wipe soap water on the air cells.If you see bubbles on the air cells, there is an air
leakage. If there is a leakage, replace the air cell.
CHECK INTERNAL TUBING
TOOLS REQUIRED:
● Phillips screwdriver for M3 screws
PROCEDURE:
1.
Remove the 8 rubber stops. Using a Philips screwdriver, remove the 8 screws that hold the rear case.
Remove the rear case.
2.
Inspect for the tubing connections. Make sure that each tube is attached to its connector.