Service
9.
Push (or pull) the loading piece of the syringe until the pressure meter shows and holds 10 mmHg, and
then press “A”. There will be a long beep.
10. Push (or pull) the loading piece of the syringe until the pressure meter shows and holds 33 mmHg, and
then press “A”. There will be a long beep.
11. Push (or pull) the loading piece of the syringe until the pressure meter shows and holds 63 mmHg, and
then press “A”. There will be a long beep. After about 5 seconds, another long beep (total 2 beeps). The
calibration is completed.
12. Switch off the power.
13. Remove the calibration tubing and put the sensor tube on senor.
14. Close the rear case and tighten all screws.
Note:
Once you displace the main PCBA, you must complete the pressure calibration and pressure
test.