Service Manual

EOLE / EOLE DC Pump

REF

 2870

REF

 2871

Summary of Contents for EOLE DC 32"

Page 1: ...2017 10 A 0 836004 5210 V1 1 www stryker com Service Manual EOLE EOLE DC Pump REF 2870 REF 2871...

Page 2: ......

Page 3: ...ment Parts 2 Service 3 Pumping set replacement 3 Air cell replacement 5 Check internal tubing 5 Fuse replacement 6 Main PCBA replacement 6 Timing motor module replacement 8 Micro switch replacement 9...

Page 4: ...ing Forget to switch off power button Switch off power switch The buzzer wire is not connected to the correct position Check the wire connections on front panel and on main PCBA Troubleshooting Quick...

Page 5: ...the 8 rubbers and remove 8 screws on the rear case Take off the rear case 2 Remove 6 screws on the base 3 Remove wire connections to power switch and front panel PCBA Take out the module set from the...

Page 6: ...pumping set and hold on this position and then tighten the screw 2 Press the vibration rods to make sure they will not touch the front object 3 Apply threadlocker around the screw to set for the best...

Page 7: ...air cells there is an air leakage If there is a leakage replace the air cell CHECK INTERNAL TUBING TOOLS REQUIRED Phillips screwdriver for M3 screws PROCEDURE 1 Remove the 8 rubber stops Using a Phili...

Page 8: ...Remove the rear case 2 Push the fuse holder together with turning counterclockwise to open it 3 Replace the fuse and discard the old one MAIN PCBA REPLACEMENT TOOLS REQUIRED Phillips screwdriver for M...

Page 9: ...e connections on the main PCBA are as below 7 After replacement perform pressure calibartion Refer to Pressure Calibration section 4 Remove the 2 screws and all wire connections on the main PCBA 3 Rem...

Page 10: ...ber stops Using a Philips screwdriver remove the 8 screws that hold the rear case Remove the rear case 2 Remove the tubes and wire connections from timing motor module 3 Remove the 2 screws on the mot...

Page 11: ...screws PROCEDURE 1 Remove the 8 rubber stops Using a Philips screwdriver remove the 8 screws that hold the rear case Remove the rear case 2 Remove the wire connections from micro switch 3 Remove the...

Page 12: ...the 8 rubber stops Using a Philips screwdriver remove the 8 screws that hold the rear case Remove the rear case 2 Remove the tubes and wire connections from front panel PCBA 3 Remove the 5 screws on...

Page 13: ...bers and unscrew 8 screws on the rear case Remove the rear case 2 Connect the deivce to power supply and switch on power switch 3 In order press A B C A then it goes into calibration mode 4 For timing...

Page 14: ...press A There will be a long beep 11 Push or pull the loading piece of the syringe until the pressure meter shows and holds 63 mmHg and then press A There will be a long beep After about 5 seconds ano...

Page 15: ...OCEDURE 1 Connect the pressure test tool to the pump Photo is for reference only 2 Switch on power 3 After inflation is completed change mode to Static 4 Choose leve1 and after the pressure is stable...

Page 16: ...2017 10 836004 5210 V1 1 www stryker com Stryker Medical 3800 E Centre Avenue Portage Michigan 49002 USA Stryker European Operations B V Herikerbergweg 110 Amsterdam 1101 CM Netherlands...