Operations Manual

ComfortGel™ support surface

2850

Summary of Contents for ComfortGel 2850

Page 1: ...2017 08 A 0 2850 109 001 REV A www stryker com Operations Manual ComfortGel support surface 2850 ...

Page 2: ...sample text ...

Page 3: ...uctions for use General warning Caution Catalogue number Serial number Batch code Manufacturer Safe working load Mass of equipment Wash by hand Do not tumble dry Do not dry clean Do not iron Allow to completely air dry Chlorinated bleach For US Patents see www stryker com patents ...

Page 4: ...sample text ...

Page 5: ...lling the support surface 7 Transferring a patient from one patient support platform to another 7 Managing incontinence and drainage 8 Selecting the appropriate CPR protocol 8 Cleaning 9 Disinfecting 10 Preventive maintenance 11 Checking the compression set of the foam 11 Quick reference replacement parts 12 Cover replacement 12 Fire barrier replacement 13 Warranty 14 Warranty exclusion and damage...

Page 6: ...and there may be an increased risk of falls when siderails are not present Serious injury or death can result from the use potential entrapment or non use potential patient falls of siderails or other restraints Consider local policies regarding the use of siderails The physician operator or responsible parties should determine whether and how to use siderails based on each patient s individual ne...

Page 7: ...erly If you do not properly rinse and dry the product a corrosive residue may be left on the surface of the product that could cause premature degradation of critical components Failure to follow these cleaning instructions may void your warranty Do not allow fluid to pool on the support surface CAUTION Improper usage of the product can cause injury to the patient or operator Operate the product o...

Page 8: ...ssure injury stages including stages 1 2 3 4 Unstageable Pressure Injury and Deep Tissue Pressure Injury and is recommended to be implemented in combination with clinical evaluation of risk factors and skin assessments made by a health care professional ComfortGel model 84 x 35 x 7 is for use on Stryker and other flat deck frames that support 84 in x 35 in support surfaces ComfortGel model 80 x 35...

Page 9: ... in x 7 in 213 4 cm x 88 9 cm x 17 8 cm Safe working load 500 lb 226 8 kg Product compliance Without fire barrier With fire barrier Compliance reference 16 CFR 1632 EN 597 1 EN 597 2 Canada Method 27 7 1979 of CAN 2 4 2 M77 16 CFR 1632 16 CFR 1633 CAL TB 129 Canada Method 27 7 1979 of CAN 2 4 2 M77 BOSTON BFD IX 11 UNI 9175 Class 1 IM EN 597 1 EN 597 2 BS 6807 Environmental conditions Operation St...

Page 10: ...ge MI 49002 USA To view your operations or maintenance manual online see https techweb stryker com Have the serial number A of your Stryker product available when calling Stryker Customer Service or Technical Support Include the serial number in all written communication Serial number location A Date of manufacture The year of manufacture is the first four digits of the serial number ...

Page 11: ...tifacts and distort readings Do not use the support surface on a larger or smaller bed frame width or length as stated in the intended use to avoid the risk of the support surface sliding and patient injury Do not use the support surface when gaps are present The risk of entrapment can develop when the support surface is placed on bed frames that leave gaps of even a few inches between the support...

Page 12: ...one patient support platform to another Always make sure that the opposite siderail is raised when placing a patient on the support surface to reduce the risk of patient fall To transfer the patient from one patient support surface to another Prerequisite Follow hospital protocols required to transfer a patient from one surface to another 1 Position one patient support platform alongside the other...

Page 13: ... If you do not properly rinse and dry the product a corrosive residue may be left on the surface of the product that could cause premature degradation of critical components Failure to follow these cleaning instructions may void your warranty CAUTION Do not allow liquid to seep into the zipper area or watershed cover barrier when cleaning the underside of the support surface Fluids allowed to come...

Page 14: ... covers to higher concentration disinfectant solutions as these may degrade the covers Do not allow liquid to seep into the zipper area or watershed cover barrier when cleaning the underside of the support surface Fluids allowed to come in contact with the zipper may leak into the support surface Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damag...

Page 15: ...hecking the compression set of the foam on page 11 Optional fire barrier for rips cracks or other visible signs of damage If excessive wear of the fire barrier is observed replacement is recommended Product serial number Completed by Date Checking the compression set of the foam Tools required 35 Straight Edge Tape Measure Procedure 1 Raise the bed height to the full up position 2 Lower the fowler...

Page 16: ...rranty information if applicable Note If the measurement is greater than 1 25 3 2 cm the compression set is out of tolerance and the support surface should be replaced See the warranty page for replacement requirements Call Customer Service at 1 800 327 0770 with the information recorded above for a replacement support surface 9 Reverse steps to reinstall 10 Verify proper orientation of the produc...

Page 17: ...ration before returning the product to service Fire barrier replacement Tools Required None Procedure 1 Raise the bed height to the full up position 2 Lower the fowler and gatch sections to the full down positions 3 Unzip the cover Start at the foot end patient right corner of the support surface and stop at the head end patient right corner 4 Fold the top of the cover to the patient s right side ...

Page 18: ...n set of less than 1 25 3 2 cm will not be replaced Warranty exclusion and damage limitations The express warranty set forth herein is the only warranty applicable to the product Any and all other warranties whether express or implied including any implied warranty of merchantability or fitness for a particular purpose are expressly excluded by Stryker In no event shall Stryker be liable for incid...

Page 19: ... 109 001 REV A 15 Warranty International warranty clause This warranty reflects U S domestic policy Warranty outside the U S may vary by country Contact your local Stryker Medical representative for additional information ...

Page 20: ......

Page 21: ......

Page 22: ...2017 08 2850 109 001 REV A www stryker com Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA ...