660 Zoom Stretcher 0660
−
009
−
002 REV
E
page 2
−
1
Checklist
All fasteners secure
Siderails move and latch properly
Engage brake pedal and push on the stretcher to ensure all casters lock securely
All casters secure and swiveling properly
Fowler operating and latching properly
Trendelenburg/Reverse Trendelenburg operating properly
Ground chain intact
No leaks at hydraulic connections
Hydraulic jacks holding properly
Hydraulic drop rate set properly
Hydraulic oil level sufficient
Lubricate where required
Engage drive wheel and ensure it is operating properly
Motion release switches working properly
Confirm battery powered functionality
Power cord not frayed
No cables worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 100 milliohms
Current leakage not more than 300 microamps
Body restraints intact and working properly
I.V. pole intact and operating properly
Oxygen bottle holder intact
No rips or cracks in mattress cover
Accessories and mounting hardware in good condition and working properly
Serial No.:
Completed By:_________________________________ Date:_____________
NOTE
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance pro-
gram should be established for all Stryker Medical equipment. Preventative maintenance may need to be
performed more frequently based on the usage level of the product.
Summary of Contents for Atlas 660Z
Page 5: ...Notes ...
Page 13: ...660 Zoom Stretcher 0660 009 002 REV E page 1 8 Notes ...
Page 25: ...660 Zoom Stretcher 0660 009 002 REV E page 3 10 Notes ...
Page 27: ...660 Zoom Stretcher 0660 009 002 REV E page 4 2 Notes ...
Page 58: ...660 Zoom Stretcher 0660 009 002 REV E page 5 31 Litter Assembly HEAD END ...