Setting cot load height with jog function
The adjustable cot load height with jog function allows you to preset the height of the load wheels to meet the ambulance
deck height up to 36 in. The jog function assists operators when loading on an incline by jogging past the preset load
height. You can set the cot load height from 26 in. to 36 in. (66 cm to 91 cm) as measured from the ground to the bottom
of the load wheel.
To set the cot load height:
1.
Locate the sensor housing on the patient right side of the cot.
2.
Using a T27 Torx driver, remove the sensor housing cover by loosening the two screws (one on each end).
3.
Adjust the left height sensor only inside the sensor housing: (
Figure 12 on page 27
):
a.
Move the sensor toward the head end to increase the set load height or move the sensor toward the foot end to
decrease the set load height.
b.
Press the retract (-) button to lower the cot to its lowest position, then press the extend (+) button to raise the
cot to its set load height position.
c.
Measure the cot height from the bottom of the load wheels to the floor.
Note:
Add an additional 1/2 in. (1,3 cm) to your deck height measurement to allow for variations in patient
weight or with other equipment that may be added to the cot.
d.
Repeat steps 3a and 3b until you reach the desired cot load height.
4.
Secure the height sensor cables. All cables should lie flat inside the housing between the sensors.
5.
Using a T27 Torx driver, reinstall the two screws (removed in step 2) to replace the sensor housing cover.
Figure 12: Adjust left sensor height
Checking the battery power level
Use the cot battery LED indictor to check the
SMRT
Pak power level. A charged
SMRT
Pak, in working condition,
provides up to 25 calls with a 250 pound patient (actual results may vary). The 24 VDC
Power-PRO
system and the
SMRT
Pak is rated for 2.4 amp-hours of electric energy.
WARNING
Do not remove the battery when the cot is active.
CAUTION
Always charge the battery before you place the product into service. An uncharged or depleted battery may cause poor
product performance.