13
Cleaning Recommendations
Inspection Recommendations
We recommend visual inspection after
cleaning prior to sterilization. Refer to the
Cleaning and Sterilization Guide.
We recommend manual precleaning and
machine cleaning method. The Navigated
MIS Jig has cleaning group IV. Refer to the
Cleaning and Sterilization Guide.
Sterilization Recommendations
To obtain optimum performance and help
prevent product damage, it is essential
that the following sterilization procedure
is followed for the Navigated MIS Jig. For
the pins, refer to the Instructions For Use
supplied with the pins.
Pre Vac
- 3 Pre-vacuumed sterilizer
cycles
- 270-273° F (132-134° C)
- Double wrapping or double
packed in sterilization
pouches
- 4 min minimum exposure
- 10 min minimum drying time
270° F
- Gravity Displacement Sterilizer
Gravity
- 270-273° F (132-134° C)
- Double wrapping or double
packed in sterilization
pouches
- 35 min minimum exposure
- 10 min minimum drying time
Flash
- Gravity Displacement Sterilizer
Gravity
- 270-273° F (132-134° C)
- Unwrapped
- 10 min preheating time
- 10 min minimum exposure
- 10 min minimum drying time
The Sterilization Recommendation‘s vali-
dation is based on ANSI/AAMI/ISO 11134
protocol.The respective individual laws,
standards, directives and instructions of
each country must be followed and have
priority of the specified procedure
(e.g.
flash gravity). It remains to the user to con
-
sider this and to acquire the corresponding
information.