![Stryker 4208 Instructions For Use Manual Download Page 10](http://html.mh-extra.com/html/stryker/4208/4208_instructions-for-use-manual_1386101010.webp)
9
Cleaning Recommendations
WARNINGS:
• Handpieces and batteries must be cleaned and steril-
ized before first and every use.
• DO NOT use solvents, lubricants, or other chemicals
unless otherwise specified.
CAUTION:
DO NOT immerse the handpiece or battery pack.
Battery Packs and Accessories
Refer to the
Cleaning Recommendations
for the battery
packs, battery pack adaptors and chargers provided
with those products.
Handpiece
1. Remove the battery pack and blade from the
handpiece.
2. Using a stiff bristle brush and mild detergent (hospital
enzymatic cleaner), scrub the debris from the
handpiece. Pay special attention to crevices and
other hard to reach areas such as seams, joints, and
details around the blade retainer and trigger areas.
CAUTION:
DO NOT allow water to run directly into the battery
contact area.
3. Rinse instruments under running water. Hold the
handpiece upright to prevent water from running into
the battery contact area.
4. If water leaks into the handpiece, tip the handpiece
back as shown below to allow drainage from a small
opening in the battery contact area.
5. Visually inspect the handpiece for any remaining de-
bris; if any is present, repeat the cleaning and rinsing
procedure using fresh hospital enzymatic cleaner.
6. Dry the handpiece with a lint-free towel.
7. After cleaning, sterilize as directed. See
Sterilization
Recommendations
.
Drain Water From Handpiece