Preventive maintenance
At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may
need to perform preventive maintenance checks more frequently based on your level of product usage.
Note:
Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
Zipper and covers are free of tears, cuts, holes or other openings
Fully unzip the cover to inspect the inside surface and core for signs of staining due to fluid ingress or
contamination
Pendant controls function properly
Max Inflate functions properly
Turn Assist functions properly
Left and Right CPR releases function properly
All electrical connections function properly
Power cord and plug are free of damage
Air cells are free of excessive wear, cracks, tears, or other damage
Fire barrier cover for excessive wear
All connectors are free of damage
Notes
•
It is recommended to replace cells every six months.
•
Replace worn or damaged components as necessary.
Product Serial Number:
Completed by:
Date:
Top cover replacement
Tools Required:
None
Procedure:
1.
Unplug the support surface power cord from the power source.
2.
Unzip the support surface top cover and discard.
3.
Replace with new top cover and zip all sides.
4.
Plug the support surface into a power source.
5.
Verify proper operation of the support surface before returning it to service.
Fire barrier replacement
Tools Required:
•
Diagonal pliers
•
Utility knife
•
Zip tie gun
Procedure: