7
Maintenance, Inspection, and Testing
• Inspect the device on a continual basis. If a problem is
observed or suspected, the device should be returned
for repair.
• Inspect all components for cleanliness. If fluid or tissue
buildup is present, repeat the above cleaning and
disinfection procedures.
Packaging
N/A
Sterilization
After performing the cleaning instructions specified above, perform
one of the following sterilization cycles.
Steam
Note for United States users: For all autoclave-compatible devices,
Stryker recommends using steam sterilization instead of liquid
chemical sterilization.
• Steam sterilization is intended only for couplers marked
AUTOCLAVE
.
• Rapid cooling, or “quenching,” the coupler after autoclaving
will result in product damage.
• The water used in the autoclave process must meet standards
for clean steam per AAMI ST 79 Appendix M – Steam Quality
requirements.
Sterilizer
Type
“Flash”
Gravity
“Flash”
Pre-vacuum
Gravity
Pre-vacuum
Min.
Temp.
132 – 137°C
(270 – 279°F)
132 – 137°C
(270 – 279°F)
132 – 137°C
(270 – 279°F)
132 – 137°C
(270 – 279°F)
Min. Cycle
Time
10 minutes
3 minutes
10 minutes
3 minutes
Product
Config-
uration
Unwrapped
Unwrapped
Double
wrapped
Double
wrapped
Drying
Time
—
—
8 minutes
8 minutes