Service Information
15
ADJUSTABLE PRESSURE COMPENSATED (P.C.) VALVE REPLACEMENT (CONTINUED)
8.
Install the replacement P.C. valve. Moisten the O–ring seal with hydraulic fluid to ensure a tight seal.
9.
Tighten the valve manually and then an additional 1/8–1/4 turn with a 13/16” wrench. Do not over –tight-
en or damage may occur to the O–ring seal.
10. Pump up the jack to maximum height and press the release pedal to lower it to check the jack’s operation.
11. Check for any hydraulic fluid leaks before replacing the base hood and litter.
NOTE
The jack descent rate was preset at the factory to lower the foot end 3–7 seconds faster than the head. It
is recommended to have the foot end lower faster to avoid patient disorientation.