Prepamatic-2
Instruction Manual
18
When using a specimen holder, 3 specimens need to be prepared at
a time. These specimens should have approximately the same size
and hardness to avoid uneven grinding.
Place the specimen holder in the levelling device and tighten the
screw on the side to fix the holder in position.
Insert the specimens into the openings and clamp them by
tightening the screw.
Insert the head with the specimen mover plate in
Prepamatic-2.
Press START
and the process will be carried out
automatically.
Using the Specimen Holder
IMPORTANT
Select screws of the correct length. Screw that are too long may stick out
from the specimen holder and result in collisions with parts of Prepamatic-2
during operation.
Starting the Process
Summary of Contents for Prepamatic-2
Page 1: ...Manual No 15347001 Date of Release 30 07 2009 Prepamatic 2 Instruction Manual...
Page 4: ......
Page 24: ...Prepamatic 2 Instruction Manual 20...
Page 54: ...Prepamatic 2 Gebrauchsanweisung Handbuch Nr 15347001 Auslieferungsdatum 30 07 2009...
Page 57: ......
Page 77: ......
Page 107: ...Prepamatic 2 Mode d emploi Mode d emploi no 15347001 Date de parution 30 07 2009...
Page 110: ......
Page 159: ...Manual No 15347001 Date of Release 0 200 Prepamatic 2 Spare Parts and Diagrams...
Page 161: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 183: ......
Page 189: ......
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 207: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......