Accutom-50
Instruction Manual
29
Sample rotation can be an advantage to get a more uniform scratch
pattern on the surface and a better planeness. On Accutom-50, three
different settings are possible.
The sample does not rotate.
The sample is rotating around its centre.
Press F3 to test sample rotation with the set parameters.
The sample is rocking/oscillating around its centre.
The angle can be adjusted from 10° to 400°.
Press F3 to test sample rotation with the set parameters.
On Accutom-50 two different ways of removing a certain amount of
material can be used.
If you know the exact position of a component underneath the
sample surface, for example 0.125 mm, this distance can be pre-set.
Move the sample as close to the cup wheel as possible, without
getting in contact with each other using the POSITION
key.
Press the POSITION
key to move the sample slowly against the
cup wheel just until there is just contact between sample and
wheel.
Press F1:RESET to set the relative X-position to zero.
After having defined the amount of material to be removed, move
the sample slightly back from the cup wheel, in the Y-direction,
press START
, and after completing the grinding process
Accutom-50 will stop precisely at the pre-defined depth.
Rotation
Rotation: Off
Rotation: Continuous
Rotation: Oscillating
Removal Mode
Material to be removed
Summary of Contents for Accutom-50
Page 1: ...Manual No 15017001 Date of Release 15 11 2012 Accutom 50 Instruction Manual...
Page 66: ...Accutom 50 Gebrauchsanweisung Handbuch Nr 15017001 Auslieferungsdatum 15 11 2012...
Page 133: ...Accutom 50 Mode d emploi Mode d emploi no 15017001 Date de parution 15 11 2012...
Page 199: ...Manual No 15017001 Date of Release 15 11 2012 Accutom 50 Spare Parts and Diagrams...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......