Struers 05765016 Instruction Manual Download Page 1

Cooling unit pumps

Instruction Manual

Original Instructions

Doc. no.: 15767026-02_A_en-us

Date of release: 2022.12.01

Summary of Contents for 05765016

Page 1: ...Cooling unit pumps Instruction Manual Original Instructions Doc no 15767026 02_A_en us Date of release 2022 12 01 ...

Page 2: ...ntents of this manual are the property of Struers ApS Reproduction of any part of this manual without the written permission of Struers ApS is not allowed All rights reserved Struers ApS 2022 12 01 2 Cooling unit pumps ...

Page 3: ...rage 8 3 1 Transport 8 3 2 Long term storage or shipping 8 4 Installation 9 4 1 Unpack the machine 9 4 2 Check the packing list 9 5 Power supply 9 6 The recirculation pump 10 7 Maintenance and service 10 7 1 General cleaning 11 7 2 Service and repair 11 7 3 Disposal 12 8 Technical data 13 8 1 Technical data 13 8 2 Legal and regulatory information 13 9 Manufacturer 14 Declaration of Incorporation o...

Page 4: ...ialographic laboratory The machine must be operated only by skilled trained personnel Do not use the machine for the following Pumping of any type of explosive and or flammable material or materials which are not stable during machining heating or pressure Furthermore the machine must not be used with consumables cutting fluids that are not compliant with function and materials of the pump Model C...

Page 5: ...uid Note that the cooling fluid can be hot and must therefore be handled with care 6 Always wear protective gloves and safety goggles when you clean and fill up the tank 7 The recirculation pump must be disconnected from the electrical power supply before it can be removed from the cooling unit 8 All safety functions must be intact and in working order If they are not they must be replaced or repa...

Page 6: ... uses the following signs to indicate potential hazards ELECTRICAL HAZARD This sign indicates an electrical hazard which if not avoided will result in death or serious injury DANGER This sign indicates a hazard with a high level of risk which if not avoided will result in death or serious injury WARNING This sign indicates a hazard with a medium level of risk which if not avoided could result in d...

Page 7: ...trical power cable before you dismantle the machine or install additional components ELECTRICAL HAZARD Switch off the electrical power supply before installing electrical equipment The machine must be earthed grounded Make sure that the actual electrical power supply voltage corresponds to the voltage stated on the type plate of the machine Incorrect voltage can damage the electrical circuit WARNI...

Page 8: ...inlet and the water outlet 3 2 Long term storage or shipping Note We recommend that you keep all original packaging and fittings for future use Disconnect the control unit from the electrical power supply Disconnect the water inlet and the water outlet Clean the machine and all accessories thoroughly Remove the control unit the pump and the water level gauge Place the items in a box Place the box ...

Page 9: ...zed hose 1 Instruction Manual set The pump is available in the following variants Small pump 05766116 05766123 05766216 05766122 05766124 Large pump 05766016 05766023 05766022 05766024 Large pump long 05765016 05765023 05765022 05765024 5 Power supply ELECTRICAL HAZARD The machine must be earthed grounded Switch off the electrical power supply before installing electrical equipment Make sure that ...

Page 10: ...he section Assembling the cooling unit in the Cooli System Instruction Manual Water level gauge Mount the water level indicator in the hole See the section Assembling the cooling unit in the Cooli System Instruction Manual 7 Maintenance and service Proper maintenance is required to achieve the maximum up time and operating lifetime of the machine Maintenance is important in ensuring continued safe...

Page 11: ...r after every 1500 hours of use When the machine is started up the display shows information about total operation time and the machines service information After 1500 hours of operation time the display will show a message reminding the user that a service check should be scheduled Note Service must only be performed by a Struers engineer or a qualified technician electromechanical electronic mec...

Page 12: ...swarf and may not be disposed of into a main drain Cooling fluid must be disposed of in compliance with local safety regulations Depending on which metals are being cut it is possible that the combination of the metallic swarf cutting debris from metals with a large difference in electrode positivity a large distance apart in the electrochemical series could result in exothermic reactions when fav...

Page 13: ... 100 120 V 60 Hz 1 x 220 240 V 50 Hz 1 x 220 240 V 60 Hz Power consumption 550 W 550 W 550 W 550 W Current nominal load 8 6 A 8 5 A 4 6 A 4 5 A Current maximum load 17 2 A 17 A 9 2 A 9 A 8 2 Legal and regulatory information FCC notice This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to pro...

Page 14: ...to which the receiver is connected 9 Manufacturer Struers ApS Pederstrupvej 84 DK 2750 Ballerup Denmark Telephone 45 44 600 800 Fax 45 44 600 801 www struers com Responsibility of the manufacturer The following restrictions should be observed as violation of the restrictions may cause cancellation of Struers legal obligations The manufacturer assumes no responsibility for errors in the text and or...

Page 15: ...16 05766906 05766905 05766929 05766930 05766931 The above listed machinery is only intended to be used with And shall not be put into service until the final machinery into which it is to be incorporated has been declared in conformity with this regulation where appropriate Struers machinery cooling units and consumables Serial no Module H according to global approach EU We declare that the produc...

Page 16: ...ännökset osoitteesta fr Pour les traductions voir hr Za prijevode idite na hu A fordítások itt érhetők el it Per le traduzioni consultare ja 翻訳については lt Vertimai patalpinti lv Tulkojumus skatīt nl Voor vertalingen zie no For oversettelser se pl Aby znaleźć tłumaczenia sprawdź pt Consulte as traduções disponíveis em ro Pentru traduceri consultați se För översättningar besök sk Preklady sú dostupné n...