background image

Page iv

PocketMIKE Operating Manual

5. Use of Couplants

Operators must be familiar with the use of ultrasonic couplants. Testing
skills must be developed so that couplant is used and applied in a
consistent manner to minimize variations in couplant layer thickness
and errors in test results. Calibration and actual testing should be
performed under similar coupling conditions, using a minimum amount
of couplant and applying consistent pressure on the transducer.

6. Doubling

Ultrasonic thickness gauges will, under certain conditions, display
readings which are twice (or, in some cases, three times) the actual
material thickness being measured. This effect, commonly known as
“doubling,” can occur below the minimum specified range of the
transducer. If the transducer being used is worn, doubling is possible at
a thickness greater than the minimum of the specified range.

When using a new transducer, any reading which is less than twice the
minimum specified range of the transducer may be a “doubled” read-
ing, and the thickness of the material being tested should be verified by
the use of other methods. If the transducer shows any sign of wear,
doubling may occur at a thickness greater than twice the minimum of
the specified range. This thickness should be determined by calibrating
the PocketMIKE on reference blocks that represent the complete range
of possible thicknesses that may be encountered in testing. This is
particularly important when the test piece is being ultrasonically
measured for the first time or in any case where the history of thickness
of the test specimen is unknown.

Physics of Ultrasound

These instruments operate on the ultrasonic pulse-echo principle,
similar to sonar. A short ultrasonic pulse is transmitted into the part by a
probe (transducer). The pulse travels through the material under test
until it encounters an interface, that is a material with substantially
different physical characteristics, such as air or liquid, at the back
surface of the part. At the interface, the pulse is reflected back to the
probe.

Page iv

PocketMIKE Operating Manual

5. Use of Couplants

Operators must be familiar with the use of ultrasonic couplants. Testing
skills must be developed so that couplant is used and applied in a
consistent manner to minimize variations in couplant layer thickness
and errors in test results. Calibration and actual testing should be
performed under similar coupling conditions, using a minimum amount
of couplant and applying consistent pressure on the transducer.

6. Doubling

Ultrasonic thickness gauges will, under certain conditions, display
readings which are twice (or, in some cases, three times) the actual
material thickness being measured. This effect, commonly known as
“doubling,” can occur below the minimum specified range of the
transducer. If the transducer being used is worn, doubling is possible at
a thickness greater than the minimum of the specified range.

When using a new transducer, any reading which is less than twice the
minimum specified range of the transducer may be a “doubled” read-
ing, and the thickness of the material being tested should be verified by
the use of other methods. If the transducer shows any sign of wear,
doubling may occur at a thickness greater than twice the minimum of
the specified range. This thickness should be determined by calibrating
the PocketMIKE on reference blocks that represent the complete range
of possible thicknesses that may be encountered in testing. This is
particularly important when the test piece is being ultrasonically
measured for the first time or in any case where the history of thickness
of the test specimen is unknown.

Physics of Ultrasound

These instruments operate on the ultrasonic pulse-echo principle,
similar to sonar. A short ultrasonic pulse is transmitted into the part by a
probe (transducer). The pulse travels through the material under test
until it encounters an interface, that is a material with substantially
different physical characteristics, such as air or liquid, at the back
surface of the part. At the interface, the pulse is reflected back to the
probe.

Summary of Contents for PocketMIKE

Page 1: ...spection Technologies LP 50 Industrial Park Road Lewistown PA 17044 Phone 1 717 242 0327 Fax 1 717 242 2606 www GEInspectionTechnologies com GE Inspection Technologies GmbH Robert Bosch Straße 3 D 50354 Hürth Phone 49 0 2233 60111 Fax 49 0 2233 601402 ...

Page 2: ......

Page 3: ...PocketMIKE Operating Manual version 1 0 ...

Page 4: ......

Page 5: ...nual provides instruction in the basic set up and operation of the GE Inspection Technologies thickness gauge There are however additional factors which affect the use of ultrasonic test equipment Specific information regarding these additional factors is beyond the scope of this manual The operator should refer to text books on the subject of ultrasonic testing for more detailed information Opera...

Page 6: ...cular materials subject to erosion or corrosion in which conditions can vary significantly in any given area should only be evaluated by fully trained and experienced operators Sound beams reflect from the first interior surface encountered Be cause of part geometry and overlapped flaws or overlapped surfaces thickness gauges may measure the distance to an internal flaw rather than to the back wal...

Page 7: ...e Probe zeroing is the process of measuring the time of flight of sound through the transducer The transducer time of flight is then automati cally removed from each measurement so that only the test piece time of flight is used to calculate and display the thickness value The probe zeroing process is performed automatically while the probe is coupled 3 Effects of Temperature on Calibration Temper...

Page 8: ... than twice the minimum specified range of the transducer may be a doubled read ing and the thickness of the material being tested should be verified by the use of other methods If the transducer shows any sign of wear doubling may occur at a thickness greater than twice the minimum of the specified range This thickness should be determined by calibrating the PocketMIKE on reference blocks that re...

Page 9: ...ded by two and multiplied by the sound velocity of the material under test The result is the thickness of the material The figure below illustrates the pulse echo principle of ultrasonic thickness measurement Dual Element Thickness Measurement Dual Probe Backwall Echo Acoustic Zero V Path of Pulse ...

Page 10: ...Page vi PocketMIKE Operating Manual ...

Page 11: ...ment Units and Resolution 13 3 5 Reversing the Display s Orientation 13 3 6 Replacing the Probe 14 3 7 Disabling Instrument Adjustment Controls 15 3 8 Rotating the Display 15 Chapter 4 Measuring Thickness 17 Chapter 5 Specifications and Declaration 19 5 1 Specifications 19 5 2 EC Declaration of Conformity 20 Chapter 6 Application Considerations 21 6 1 Material Consistency in Test Specimen 21 6 2 F...

Page 12: ...Page viii PocketMIKE Operating Manual ...

Page 13: ...soon as possible after the low battery indicator appears on the instrument s display To install the battery refer to Figure 1 1 and follow these steps Step 1 Unscrew the probe locking ring and remove the probe from its sealed position Step 2 Insert one AA battery in the instrument Make sure the battery poles are oriented as shown in Figure 1 1 The instrument will not function if the battery is ins...

Page 14: ... operation the instrument automatically powers off Instrument settings are saved and restored when batteries are replaced and the instrument is again pow ered on 1 2 Powering On the Instrument To power on the instrument simply press Pressing and holding this key for more than three seconds when the instrument is on will cause it to power off If the instrument is uncoupled and no keys are pressed f...

Page 15: ...yed in the instrument s screen and adjusted using various combinations of key presses This chapter identifies the display screen s features the keypad functions and the general display structure Topics covered include Keypad Operations section 2 1 Display Indicators section 2 2 On Screen Menu Structure section 2 3 FIGURE 2 1 PocketMIKE ...

Page 16: ...3 2 and 3 3 Changes setting when in Backlight Adjustment mode section 3 1 Reverses display orientation section 3 5 Changes on screen values when in thickness and velocity calibration modes sections 3 2 and 3 3 Changes setting when in Backlight Adjustment mode section 3 1 and Simultaneous pressing of these keys selects measurement units and resolution section 3 4 Simultaneous pressing and HOLDING t...

Page 17: ... referenced manual section FIGURE 2 2 Each of these screen indicators are displayed only under certain conditions Indicates instrument is in Thickness Measurement mode when display is not inverted chapter 4 Thickness Calibration mode is in process section 3 2 Velocity Calibration mode is in process section 3 3 Indicates instrument is in Backlight Adjustment mode section 3 1 Indicates a low battery...

Page 18: ...urn for service CAL Calibration mode for sound velocity is active Thickness may be calibrated 2 3 Types of Display Screens The PocketMike offers three general display screen modes similar to the three shown below Thickness Measurement Display Figure 2 3 Calibration Mode Figure 2 4 Backlight Adjustment Mode Figure 2 5 FIGURE 2 3 In thickness measurement mode appears along the display s upper edge O...

Page 19: ...libration display varies depending on whether velocity or thickness calibration mode is indicated See sections 3 2 and 3 3 for more details related to instrument calibration FIGURE 2 5 Pressing twice when operating in Thickness Measurement mode launches the Backlight Adjustment mode Pressing or sets backlight to On OFF or Auto section 3 1 ...

Page 20: ...Page 8 PocketMIKE Operating Manual ...

Page 21: ...ution section 3 4 Reversing the Instrument s Display Orientation section 3 5 Replacing a Probe section 3 6 Disabling All Adjustment Controls section 3 7 Rotating the Display section 3 8 NOTE Prior to setting up an instrument a battery must be installed and the instrument must be powered on Refer to Chapter 1 for information on installing a battery and turning the instrument on NOTE Before using th...

Page 22: ... one or two times to activate the Backlight Adjustment mode When this mode is active appears in the display s corner Press or to select between the following backlight modes On OFF Auto When Auto mode is selected the backlight automatically illuminates each time a key is pressed or when the probe is coupled The backlight remains illuminated for approximately 4 seconds after which it auto matically...

Page 23: ... 2 Thickness Calibration The instrument offers two calibration modes thickness and velocity section 3 3 To launch thickness calibration mode press anytime the instrument s display shows the indication Note that throughout the calibration process appears and flashes along the display s upper edge Figure 3 1 illustrates the displayed param eters and key presses required to navigate through the thick...

Page 24: ...t the velocity calibration process appears and flashes along the display s upper edge Figure 3 2 illustrates the displayed parameters and key presses required to navigate through the velocity calibration process NOTE A thickness standard is not required to perform a velocity cali bration However following a velocity calibration the instrument s ac curacy should be checked using a sample of known t...

Page 25: ...ll also determine the units in which acoustic velocity is displayed 3 5 Reversing the Display s Orientation The displayed thickness and units of measurement or can be reversed displayed upside down to allow for easy display screen reading with the instrument in either the probe up or probe down orientation To reverse the display simply press or anytime the instrument s display shows A second press...

Page 26: ...damaged probe transducer contact surfaces can affect measurement accuracy The probe and locking ring are replaced as an assembly Refer to Figure 3 3 for the probe replace ment procedure Always recalibrate the instrument after the probe is replaced or removed FIGURE 3 3 Probe Replacement ...

Page 27: ...ion units of measurement and backlight setting can not be adjusted This mode is disabled by repeat ing the simultaneous key press and is indicated when donE appears on the display Note that powering the instrument off and back on does not disable the lockout mode 3 8 Rotating the Display The plastic display housing on the top of the PocketMike can be rotated through 180 degree of motion To rotate ...

Page 28: ...Page 16 PocketMIKE Operating Manual ...

Page 29: ...lly powered off The PocketMIKE measures thickness in units of inches or mm Read the following notices and instructions before measuring thickness When is displayed the instrument is ready to measure thick ness Follow these steps to measure thickness Step 1 Remove dirt loose material and couplant residue from the surface of the test piece Step 2 Be sure that the instrument has been calibrated to ma...

Page 30: ...in the user specified measurement unit inches or millimeters and resolution as described in section 3 4 When the probe is uncoupled will no longer appear along the edge of the display but the instrument will continue to display the last measured thickness Refer to Figure 4 1 to interpret display indicators FIGURE 4 1 Thickness Measurement Mode NOTE Only couplants approved by GE Inspection Technolo...

Page 31: ...FF On Approximate Measuring Range 0 040 1mm to 9 999 250 mm Material and Application Dependent Display Resolution 0 001 inch for 10 000 inches 0 01 inch for 10 00 inches 0 01 mm for 100 00 mm 0 1 mm for 100 00 mm Reading Stability 001 inches Operating Temperature 10 C to 50 C 10 F to 120 F Storage Temperature 20 C to 60 C 4 F to 140 F Probe Surface Temperature 10 C to 100 C 10 F to 212 F Maximum c...

Page 32: ...mity with the requirements of the directive 89 336 EEC is proved by meeting the standard specifications IEC 61000 6 2 2001 Immunity to interference for industrial environment IEC CISPR11 1998 A1 1999 A2 2002 Class A Group 2 Emitted interference for ISM equipment Note Class A instruments are instruments suitable for use in all other areas except for the living area and except for areas which are di...

Page 33: ... measurement accuracy will be obtained if the work piece and the calibration piece are the same temperature shape material and size NOTICE Since sound velocity varies from test piece to work piece accuracy of measurement depends on consistency of sound velocity Sound velocity also varies with internal stress so that heat treatment of the material will also effect accuracy The following table expre...

Page 34: ...ondition The problem can usually be corrected by rotating the probe so that the crosstalk barrier is at a right angle to the grooves It is possible for the surface of a test piece to be too rough to permit a good reading Excess couplant could be trapped between the probe and surface under test causing a false reading A very rough surface may prevent coupling altogether no coupling indicator This p...

Page 35: ...age 23 FIGURE 6 1 Crosstalk Barrier Orientation PROBE CONTACT FÄCE CROSSTALK BARRIER LONG AXIS OF TUBE When using a flat dual probe position the crosstalk barrier at a right angle to the long axis of the part as shown in Figure 6 1 ...

Page 36: ...Page 24 PocketMIKE Operating Manual ...

Page 37: ...PocketMIKE Operating Manual Page 25 Troubleshooting Guide Should your instrument malfunction refer to the guidelines that correspond to the problem you re experiencing ...

Page 38: ...Page 26 PocketMIKE Operating Manual ...

Page 39: ...spection Technologies GE Inspection Technologies warrants new transducers manufactured by GE Inspection Technologies and delivered to the original retail purchaser F O B GE Inspection Technologies s factory to be free from defects in material and workmanship under normal use and service for ninety 90 days from the date of purchase These warranties are subject to the following limitations to which ...

Page 40: ...een subject to proper care and maintenance GE Inspection Technologies shall not be liable for any damages whether direct or indirect economic commercial incidental or consequential and whether arising from GE Inspection Technologies s negligence breach of contract product liability warranty or any other reason To any instruments or transducer not manufactured by GE Inspection Technologies For equi...

Page 41: ...ly provide the following information Company Name Contact Name Phone Number Company Address Model Number Serial Number Accessories Description of problem or work required Any additional comments and send it with your unit to GE Inspection Technologies LP 50 Industrial Park Road Lewistown PA 17044 U S A Phone 1 717 242 0327 Fax 1 717 242 2606 or GE Inspection Technologies GmbH Robert Bosch Straße 3...

Page 42: ...it from transit damage Static sensitive parts must be packaged in anti static bags foam or tubes All products should be sent back in their original carry cases or wrapped in bubble wrap or other available packaging material GE Inspection Technologies warrants all repairs for a full 90 days ...

Page 43: ...c r o P 0 0 2 2 0 0 6 5 d l o G 0 0 3 1 0 0 3 3 l y t u B r e b b u R 0 3 7 0 0 0 9 1 l e n o c n I 0 0 2 2 0 0 6 5 c l u V r e b b u R 0 0 9 0 0 0 3 2 d a e L 0 5 8 0 0 0 2 2 r e v l i S 0 0 4 1 0 0 6 3 m u i s e n g a M 0 0 3 2 0 0 8 5 l e e t S 0 0 3 2 0 0 8 5 e s e n a g n a M 0 0 8 1 0 0 6 4 n i T 0 0 3 1 0 0 3 3 m u n e d b y l o M 0 0 5 2 0 0 3 6 m u i n a t i T 0 0 4 2 0 0 1 6 l e n o M 0 ...

Page 44: ...Page 32 PocketMIKE Operating Manual ...

Page 45: ...dex A Auto 10 B Backlight 10 Backlight Adjustment 10 Batteries 1 Battery indicator 2 Brightness 10 C CAL 6 Calibration 11 Curved Surfaces 22 D Display 5 donE 6 E EC Declaration of Conformity 20 F FAIL 6 Features 19 Flaws in the Specimen 22 ...

Page 46: ...4 L Lockout 15 M Material Consistency 21 P Powering On the Instrument 2 Probe transducer 14 R Repair 29 Resolution 13 Reverse the display 13 S SAFE 15 Service 29 Specifications 19 Surface 22 T Temperature 17 Thickness 17 Thickness calibration 11 Troubleshooting 25 ...

Page 47: ...PocketMIKE Operating Manual Page 35 U Units of measurement 13 V Velocity calibration 12 W Warranty 27 ...

Reviews: