background image

Turning on and shutting down

 the scanner

Turn the scanner on by pressing the scanner button. The indicator light at the back of the scanner will turn blue. 
After a short while, the scanner will connect to the MOVE+/POD and the indicator light will turn green.

Action desired

What to do

Notes

Wake up the scanner

Press the scanner button

The Indicator light at back of scanner 
turns on (blue, then green when connected to PC)

Turn the scanner off

Press the shut-down button  
on the user interface or take out 
the battery.

Click the scanner icon in the top right corner and the 
shut-down button will appear on the screen, or follow 
the menu bar steps: More > Settings > TRIOS > Scanner 
Management > Turn off scanner

Automatic shutdown

 of the scanner 

The scanner will shut down automatically after approx. 10 minutes when not connected to a 3Shape TRIOS

®

 system i.e.,  

the status light at the back of the scanner is blue. If the scanner shuts down unintentionally, you can turn it back on 
by pressing the scanner’s button. The scanner will reconnect to the PC after approx. 10-20 seconds. Once the scanner 
has been reconnected, instant-heat technology enables the system to be scan-ready in seconds.

If the scanner is connected to a TRIOS system, the scanner will not power down in order to keep the tip warm and 
ready (i.e., the status light at the back of the scanner is green). Please note that the scanner will consume battery 
while in stand-by mode.

Inserting the WI-FI adapter

 into the PC (POD systems) 

Insert the Wi-Fi USB adapter delivered with the scanner into a free USB slot on the PC. 

The Wi-Fi adapter (that goes into the USB port) must be connected directly to one of the USB ports on the PC. Do not 
insert any form for hub or similar hardware between the adapter and the PC, as this can make the connection

 

unstable. 

Important note: 

Use the USB slot furthest away from the position of the scanner mounted in the POD (as shown in 

the picture below). If the Wi-Fi adapter is too close to the scanner (approx. 30cm/12” or less), then the wireless signal 
may b too strong and cause temporary loss of the connection to the scanner.

Make sure that you do not apply mechanical stress (bend) to the WI-FI adapter. This may accidentally happen 
when moving e.g. a laptop.

7

Summary of Contents for 3Shape TRIOS 4

Page 1: ...3Shape TRIOS Info Pack TRIOS 4 Let s change dentistry together...

Page 2: ...attery Turning on and shutting down the scanner Automatic shutdown of the scanner Inserting the WI FI adapter into the PC POD systems General recommendations for a stable wireless connection Wireless...

Page 3: ...pending on battery charge level when charging is initiated What is the battery lifetime Approximately 80 after 300 cycles and 70 after 500 cycles How long does it take to connect a wireless scanner to...

Page 4: ...for detailed instructions Wired setup Connecting the smart cord to TRIOS 4 wired setup Step 1 Make sure that the smart cord edge is aligned as shown in the picture above Step 2 Gently press the smart...

Page 5: ...tothepod Thecable goes into the left connection socket on the back side of the pod the side that has two connection sockets If you need to release the Ethernet cable press on the small tab on the top...

Page 6: ...w scanner field and then press the green plus icon button Hint The serial number is located on the front tube It is the number written in the largest font starting with 1HC Step 8 The indicator light...

Page 7: ...PC 3 Follow the steps Starting up for the first time to use the scanner on the second PC If the wired setup has previously been used on the second PC 4 Shut down the scanner take the POD and smart co...

Page 8: ...correctly Attempting to insert incorrectly may cause permanent damage to the scanner Please refer to the Safety Setup Guide for full instructions on how to install the batteries The information below...

Page 9: ...l reconnect to the PC after approx 10 20 seconds Once the scanner has been reconnected instant heat technology enables the system to be scan ready in seconds If the scanner is connected to a TRIOS sys...

Page 10: ...be possible to remove the cause of critical the noise In such cases a wired scanner may be a better choice for your practice Many clinics have x ray protective measures installed in the walls This wil...

Page 11: ...e for thermal variations or aging effects due to wear For a detailed description of the calibration process please refer to the TRIOS Safety and Setup guide If a connection to the scanner is lost If t...

Page 12: ...I adapter and a copy license dongle are inserted in the second PC 3 Follow the steps Starting up for the first time to use the scanner on the second PC If the wireless setup scanner has previously bee...

Page 13: ...ate Blue Scanner is on and attempting to connect Yellow Battery is 5 or lower Change the battery now The scanner will continue to scan until the battery is completely exhausted Green Scanner is on con...

Page 14: ...ng Steady green The battery is fully charged Charging begins automatically when a battery s temperature is between 0 and 45 degrees Celsius During charging the battery s temperature can go up to 52 de...

Page 15: ...ll consume battery if it is turned on and connected to the PC MOVE Changing batteries when power runs out If the battery runs out simply remove the discharged battery and replace it with a fully charg...

Page 16: ...arge half of the battery capacity each time before recharging you will consequently have approximately 1000 of these half cycles Smart tips Smart tip for 3D color scans The 3Shape TRIOS 4 tips use a t...

Page 17: ...ch smart tip has an integrated mirror The optical system that is responsible for the scan quality must be calibrated to each mirror of separate smart tips For a more detailed description of the color...

Page 18: ...Wi Fi adapter into the USB port Battery level has dropped to 5 or less Tech Specs Please refer to the product catalog which may be found on the partner portal partner 3shape com If you don t have a l...

Page 19: ...ns for both dental practices and labs Two graduate students founded 3Shape in Denmark s capital in the year 2000 Today 3Shape has over 1 500 employees serving customers in over 100 countries from 3Sha...