![STRALUS SLLS002 User Manual Download Page 27](http://html1.mh-extra.com/html/stralus/slls002/slls002_user-manual_1383335027.webp)
Troubleshooting
6
27
Eff ect
Possible Cause
Verifi cation
Corrective Action
Bed does not stay
Braking system is not
Brake pedal is not
Brake pedal is not
in place
in Total Brake position
engaged in lowest position
engaged in lowest position
Braking System is
Brake pedal is in lowest
Contact your sales
malfunctioning
position, but bed does not
representative or a qualifi ed
stay in place
Stralus service technician
Bed does not move
Bed is not plugged in
Check if the power lead
If unplugged, plug power
up or down
and optional battery is fl at
is plugged into mains
lead into mains power.
power
Allow 3 hours to charge
battery (if fi tted)
Actuator is faulty
Contact your sales
representative or a qualifi ed
Stralus service technician
Backrest does not
Actuator cable is
Check if cables have
If loose, fully plug
move up or down
unplugged
come loose from control
cables into control box
box (under top frame at
head
end)
Actuator is faulty
Contact your sales
representative or a qualifi ed
Stralus service technician