Technical data
Rated mains voltage
120 V
Line frequency
50/60 Hz
Response value
≤ 25 kOhm
Response time
< 2 seconds
Power consumption
4 VA at 120 VAC / 50/60 Hz
Signal displays
Acoustic signal
LED green: Operation
LED yellow: Isolation error
LED yellow: Transformer limit temperature reached
Transformer power utilization:
LED green: >30 %
LED yellow: >60 %
LED red: >90 %
Pulsating with an isolation error, continuous tone with
excess temperature
Ambient operating temperature
0 °C … +40 °C
Relative humidity
30 … 75 %
Protection class
II
Degree of protection
IP 40
Dimensions of evaluation element
[H x W x D]
125 x 67 x 40 mm
Dimensions of control unit
80 x 44 x 24 mm
Observed standards
EN 60601-1, EN 60601-1-2, EN 14971
Earth leakage monitor order no.:
Contains:
Control unit, evaluation element
and test plug
UG 400
EMC limit value class:
B (as per EN 55011+A2)
Only store in closed or roofed rooms,
storage temperature/transport temperature:
-10 °C to +60 °C, humidity 5 to 95 %
Do not expose the devices to chemical vapors
or gases which are corrosive, highly flammable,
combustible, explosive or conductive.
THIS DEVICE IS NOT SUITABLE TO BE USED IN
AN OXYGEN RICH ATMOSPHERE.