19 620 02 0117
14
System Description
3.4
Transport
It requires two persons to transport the device after unpacking riskless and safe to the
treatment place.
NOTE
Make sure that your hands are dry and free of grease.
• Dismount the handpiece holder before transporting the device.
• Grip on the front and on the rear side with both hands on the bottom of the de-
vice as shown in in Fig. 3-4 and lift it carefully.
Fig. 3-4 Transporting the device
•
Set the device slantly down in order to avoid squeezing the fingers.
3.5
Installation Instructions
3.5.1
Mounting the handpiece holders
There are openings for 6 handpiece holders and /or suction cup holder - 3 on the
right side and 3 on the left side of the D-ACTOR 200 in each case in the front, in the
middle and rear section.
3.5.1.1
Handpiece holder
There are two different handpiece holder:
– for handpieces D-ACTOR
– for handpieces V-ACTOR.
1
1 Openings for insertion of the
handpiece and suction cup
holder
Fig. 3-5 Mounting the handpiece holder