Preparation
Instructions for use • C-MAC Video Laryngoscopes Series 8404 • USB826_EN_V1.2_11-2021_IFU_CE-MDR
16
3.
Ensure that the components and connections are complete, intact, and positioned
correctly.
Defect pattern: Defective socket, damaged multifunction button
4.
Check whether there are residues or moisture at the interfaces.
5.
Use a magnifying glass to check that the optics is complete, intact, and securely
positioned.
5.2.2 Functional test
Inspect the product for the following characteristics:
–
Fully functional control keys
–
Securely positioned connecting cable
–
Functional image transmission
–
Sufficient image quality
–
Light transmission
5.3 Connecting the C-MAC Monitor
1.
Insert the video connecting cable into the socket of the video laryngoscope. Pay
attention to the orientation pin on the plug and socket.