Problem
Causes and Actions
The WAA Programmer is
unresponsive: The indicator
light does not turn on. The
stimulation amplitude keys
do not respond.
The WAA Programmer is not powered on.
§
Turn the power ON.
The WAA Programmer system has “frozen”.
§
Turn the power OFF and wait 5 seconds
before turning the power back ON.
The battery is not charged.
§
Recharge the battery by placing the WAA
Programmer on the charger.
Dropped WAA Programmer:
Your WAA Programmer falls
off a cabinet or table.
WAA Programmer is designed to withstand
a short drop on a hard surface and still
operate normally. Try powering ON the
WAA Programmer and allow it to transmit
for 10 minutes while not worn on the body.
If the WAA appears to be functioning
properly, try using it.
Fluid on the WAA
Programmer: Fluid was
spilled onto the WAA
Programmer or the WAA
Programmer was dropped
into water.
WAA Programmer is not waterproof, and
water can damage the device.
§
Immediately
remove
the
WAA
Programmer from the water, then dry
the WAA Programmer with a towel
damped with clean tap water. Unplug the
power cord and allow the WAA
Programmer to dry at room temperature
for 24-48 hours.