ASTOPAD
®
Instructions for Use
42
10.5 Sensor defect alarm “A5”
(medium priority alarm condition)
Control panel
Device response
•
The display
for the affected output shows “A5”.
•
The Start LED
and the Alarm LED
flash.
•
An acoustic alarm signal is triggered.
•
The heating process is stopped at the two outputs.
Alarm condition
This alarm is triggered if one or more sensors in the
applied part are defective.
Possible reasons
►Required
action(s)
Defective sensor(s) in the applied part.
►Send the applied part to the local sales point.
Defective connection cable on the applied part.
►Send the applied part to the local sales point.
Defective cable plug on the applied part.
►Send the applied part to the local sales point.
Defective extension connection cable.
►Replace the extension connection cable.
Defective connecting socket for output A or B on the
control unit.
►Send the control unit to the local sales point.