Steris VHP MD140X Manual Download Page 7

7

EQUIPMENT DRAWINGS (REQUIRED FOR INSTALLATION)

Equipment Drawing 

Part Number

Equipment Drawing Title

387358-924

Single Door, Recessed

387358-923

Single Door, Cabinet

SINGLE DOOR, CABINET MODEL SHOWN

Dimensions are in inches [mm]

No

t F

or

 In

sta

lla

tio

n

SIDE VIEW

PLAN VIEW

FRONT VIEW

Drawing is not to scale.

Dimensions are typical.

30 1/2

[775]

DOOR SWING

Summary of Contents for VHP MD140X

Page 1: ... compatible with a broad range of materials and has rapid Biodecontamination Cycle times There are no toxic by products created by the Biodecontamination Cycle only water vapor and oxygen are produced The VHP MD140X Biodecontamination Unit performs three programmed Biodecontamination Cycles Lumen Cycle approximately 55 minutes to complete Non Lumen Cycle approximately 28 minutes to complete Flexib...

Page 2: ...er right corner Each bar represents approximately four cycles Unit is available with single door in either a freestanding or recessed configuration Unit installation requires no plumbing ventilation or air supply only a dedicated electrical connection is needed A power cord is supplied for this connection Unit is equipped with a racking system for articles to be biodecontaminated Unit is equipped ...

Page 3: ...nt and continues to evacuate for programmed time to reduce chamber vapor concentration Once Aeration phase is complete chamber pressure returns to atmospheric and chamber door is unlocked NOTE Automatic door locking mechanism keeps chamber locked during entire Biodecontamination Cycle BIODECONTAMINATION ASSURANCE PRODUCTS Vaprox 59 Hydrogen Peroxide Sterilant 59 stabilized aqueous solution of hydr...

Page 4: ...t Floor surface must be hard and level 4 Biodecontamination Unit should not be installed in an area not compatible with oxidizers Consult the MSDS regarding hydrogen peroxide sterilant 5 STERIS recommends maintaining and operating Biodecontamination Unit in area where temperature ranges from 41 104 F 5 40 C and has ventilation system exchanging area air at least 10 times per hour 6 STERIS recommen...

Page 5: ...5 ...

Page 6: ...mber Equipment Drawing Title 387358 924 Single Door Recessed 387358 923 Single Door Cabinet SINGLE DOOR RECESSED MODEL SHOWN Drawing is not to scale Dimensions are typical Dimensions are in inches mm Not For Installation SIDE VIEW PLAN VIEW FRONT VIEW 30 1 2 775 DOOR SWING ...

Page 7: ...umber Equipment Drawing Title 387358 924 Single Door Recessed 387358 923 Single Door Cabinet SINGLE DOOR CABINET MODEL SHOWN Dimensions are in inches mm Not For Installation SIDE VIEW PLAN VIEW FRONT VIEW Drawing is not to scale Dimensions are typical 30 1 2 775 DOOR SWING ...

Page 8: ...440 354 2600 800 444 9009 www STERISLifeSciences com SD980 2013 STERIS Corporation All rights reserved 03 01 13 This document is intended for the exclusive use of STERIS Customers including architects or designers Reproduction in whole or in part by any party other than a Customer is prohibited ...