High Power Vacuum Pump
is supplied to produce cycle
vacuum pulses that remove air and moisture from chamber.
Direct drive rotary vane pump is quiet (<60 dB) with low
vibration. Powerful 3 HP (2.2 kW) motor produces a
displacement of 53 CFM (90 m
3
/hr) and helps alleviate moisture
sensitivity in biodecontamination unit. Biodecontamination unit
operating pressure is from atmospheric pressure down to less
than 1 Torr.
CONTROL DESCRIPTION
The VHP MD140X Biodecontamination Unit is equipped with
an Allen-Bradley Compact Logix
™
(PanelView Plus
™
6 1000
Display
1
) control. PLC features a 10.4" color TFT display with
touch screen.
Screens are color coded for operator convenience as follows:
• Control Screens -
» Condition Phase - Green
» Biodecontamination Phase - Blue
» Aeration Phase - Violet
• Service Screens - Light Blue
• Option Screens - Purple
• Alarm Screens - Red
NOTE: This Biodecontamination Unit permits no manual
control of the Biodecontamination Cycles.
1
CompactLogix™ and PanelView™ Plus 6 1000 are trademarks of
Allen-Bradley, a Rockwell Automation Company.
CYCLE DESCRIPTION
The VHP MD140X Biodecontamination Unit is equipped with
three programmed Biodecontamination Cycles: Lumen Cycle
(55 minutes), Non Lumen Cycle (28 minutes) and Flexible
Cycle (35 minutes). Each Biodecontamination Cycle proceeds
through three phases: CONDITION, BIODECONTAMINATION
and AERATION.
Basic description of a Biodecontamination Cycle (example):
•
CONDITION
— This cycle phase is a set time vacuum
pulse to remove air and moisture from the chamber. When
setpoint is reached, load is tested for acceptable moisture
content. If content is acceptable, cycle proceeds. If not,
Condition pulse repeated.
NOTE: If Condition phase fails the second moisture check,
the cycle Aborts.
•
BIODECONTAMINATION
— This cycle phase is a series of
four pulses. Each pulse consists of: vacuum pulled to
setpoint; Vaprox 59 Hydrogen Peroxide Sterilant vapor
drawn into chamber; hold for programmed time; filtered air
is introduced to setpoint; hold for programmed time; deep
vacuum pulled to setpoint.
•
AERATION
— This cycle phase pulls a vacuum to
setpoint and continues to evacuate for programmed time
to reduce chamber vapor concentration. Once Aeration
phase is complete, chamber pressure returns to
atmospheric and chamber door is unlocked.
NOTE: Automatic door locking mechanism keeps chamber
locked during entire Biodecontamination Cycle.
BIODECONTAMINATION ASSURANCE
PRODUCTS
Vaprox 59 Hydrogen Peroxide Sterilant
– 59% stabilized
aqueous solution of hydrogen peroxide designed for use with
VHP Biodecontamination Units and Accessories (EPA Reg. No.
1043-123). Order PB033US (950 mL).
Steraffirm
®
NB305 Process Indicator
– Chemical indicator
designed for use with hydrogen peroxide vapor.
Steraffirm
®
PCC051 Process Indicator
– Chemical indicator
designed for use with hydrogen peroxide vapor.
Spordex
®
NA330 Biological Indicator
– E6
Geobacillus
stearothermophilus
inoculated stainless-steel coupon
designed for use with hydrogen peroxide vapor.
Spordex
®
NA114 Biological Indicator Media
– Culture media
designed for use with Spordex and Spordi
®
biological
indicators.
ACCESSORIES
Seismic Tie Down Kit
– Required for seismic installations for
safe operation of Biodecontamination Unit.
Trays And Organizer Clips
– Trays are lightweight, durable
and available in a variety of sizes to fit your instruments while
organizer clips anchor instruments in-place for transportation.
Tray Mats
– Mats are lightweight, durable and feature flexible
prongs which loosely grip instruments.
Pouches
– Vis-U-All™ pouches combine durable Tyvek with
a clear 2.0 mL film to offer a wide variety of pouch sizes and
styles in heat-seal, self-seal and customizable roll tubing.
PREVENTIVE MAINTENANCE
Customers are encouraged to contact STERIS Life Sciences
concerning annual maintenance programs. Preventive
maintenance, adjustments and replacement of worn parts are
provided on a scheduled basis to help ensure optimal
equipment performance and help minimize untimely and costly
interruptions. STERIS Life Sciences maintains a worldwide staff
of well-equipped, factory-trained technicians to provide these
services, as well as on-site installation, training and expert
repair services. Contact STERIS Life Sciences
2
for details.
2
(800) 444-9009 or www.sterislifesciences.com
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
REGULATIONS.