3
3-1
Component Identification
Operator Manual
920011-649
3.1 Processor –
General
Before operating the Reliance
™
EPS Processor, it is important to
become familiar with the location and function of all major
components and controls (see Figures 3-1 to 3-4).
NOTE: For a 50 Hz processor, also refer to Operator Manual
Information Supplement (P920501-322) for specific components.
3.1.1 Chamber Door
See Figure 3-1.
A front-loading, drop-down door is provided to facilitate loading of the
processor. The door is constructed of 316 stainless steel covered with
a plastic panel. The panel is strong and easy to clean. A tempered
glass window allows the operator to view the chamber interior while a
cycle is in progress. Panel and glass window remain relatively cool
to the touch.
A microswitch prevents a cycle from starting if the door is not fully
closed. The processor is also provided with an
UNLOCK
toggle
switch that must be pressed before opening the door.
3.1.2 Service Access Panel
See Figure 3-1.
The front bottom service access panel provides an easy access to
the 0.2
μm bacterial-retentive filter, cleaner injection pumps, fresh
container detection meter and other components for maintenance.
3.1.3 Base
The processor base offers an ergonomic way of loading and
unloading the endoscope processing support.
3.1.4 Bag-in-a-Box
Compartment
See Figure 3-1.
A three-shelf compartment is provided at the right of the chamber. The
top two shelves are designed to receive Bag-in-a-Box containers and
are provided with specially-designed connectors. The Bag-in-a-Box
compartment drop-down door is made of the same easy-to-clean
plastic as the chamber door panel.
COMPONENT IDENTIFICATION
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...