![Steris CO2EFFICIENT 710302 Operator'S Manual Download Page 36](http://html.mh-extra.com/html/steris/co2efficient-710302/co2efficient-710302_operators-manual_1374337036.webp)
CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR Operator’s Manual
Page 35
Condition
Possible Causes
Field Checkout
“CONTINUED”
My CO
2
supply tank is
prematurely emptying.
What could be the problem?
High Pressure Hose from
CO
2
supply tank is not
installed properly.
Inspect the CO
2
supply tank yoke and verify that the plastic
disc that creates the gas seal at the CO
2
supply tank valve is
in-place and free from defect.
After verifying all High Pressure Hose sealing surfaces are
clean and free of defect; reconnect the High Pressure Hose to
the CO
2
supply tank yoke and CO
2
EFFICIENT
ENDOSCOPIC INSUFFLATOR. Use an adjustable or
opened-ended ¾" wrench to tighten securely. DO NOT USE
ANY TEFLON TAPE OR THREAD SEALING
COMPOUNDS.
Reconnect the yoke to the CO
2
supply tank valve making
sure that the plastic sealing disc is in place.
Open the CO
2
supply tank valve. If CO
2
continues to
prematurely empty or should you hear CO
2
escaping from
any of these connections or the High Pressure Hose itself,
contact Customer Services.
Disc