EFFICIENT ENDOSCOPIC INSUFFLATOR Operator’s Manual
Page 15
Section 7.0 REAR PANEL CONTROLS
[8] GAS INPUT PORT
For connecting a “D” or “E” supply tank using the High Pressure Hose and Yoke
assembly provided.
WARNING:
Do not allow liquid CO
2
to enter the unit. This can be prevented by assuring that
the CO
2
supply tank is maintained in a vertical position at all times.
[9] AC POWER CONNECTION
Universal AC line input device – nominal AC line voltage 100 to 240 VAC
frequency 50/60 Hz. The AC line voltage should not drop below 90 VAC or
exceed 264 VAC. There are no switches or other AC line configuration
requirements.
The line cord (mains lead) supplied with this unit is designed and approved for
use in the USA and Canada only, and should not be used outside these countries.
For use outside of the USA and Canada, your Distributor will supply a line cord
that is approved for use in your country.
AC input is with a standard hospital grade line cord (mains lead). Connection
should be to hospital grade receptacles only.