32
8.3
Software/Hardware Issues
PROBLEM
POTENTIAL CAUSE
SOLUTION
Operating system fails to start.
BIOS screen error is displayed or
beeping is heard emanating from
instrument.
RAM disconnected or Operating
System software corruption
Copy any error messages
displayed and contact Technical
Support
Operating system boots up but
RoboSep™-S fails to fully initialize
Power switch (front of instrument)
not left off long enough after
shutdown. It takes up to 10 seconds
for the electronics to fully reset after
the power switch has been turned
off.
After shutdown, turn off the power
switch and count to 10 before
turning the power back on again
RoboSep™-S software application
fails to start
Ethernet cable is connected
Disconnect Ethernet cable and
restart the instrument. If issue
persists, contact Technical Support.
RoboSep™-S won’t shut down
Previous run not complete or needs
to be unloaded
Let run finish. Press unload button
on the Run Progress screen.
RoboSep™-S won’t resume a
paused run
The lid is open
Close the lid and press Resume on
the touchscreen
Cannot add protocols in the My
Protocols tab
Protocols are assigned to a quadrant
Remove protocols assigned to any
quadrant and then add protocols to
a user profile (section 4.2.1)
Touchscreen not functioning
Physical damage to screen or
recalibration problem
Contact Technical Support