5. Installation Guide

66

Initial Set Up 

The UTS mattress defines a new product category within the pressure care mattress range spectrum. A step up 
from traditional hybrid system. Allows the user to adopt and customize the system to suit a wider range of care 
requirements. 
 
1. 

STATIC COMFORT MODE

  Mattress is fully inflated and held in static mode with top surface foam in position underneath top PU outer  
  cover (mattress is plugged in and switched on in standby mode) 
 
2. 

DYNAMIC COMFORT MODE 

  Mattress is fully inflated and operated in auto dynamic mode with top surface foam in position underneath.  
  Top PU outer cover. 

3. 

DYNAMIC MODE (AUTO) 

  Mattress is fully inflated and operated in auto mode with patient in direct contact with alternating cell surface for  
  maximum reactive therapy. Foam top surface is stored inside the pocket beneath the dynamic cells. 

4. 

DYNAMIC MODE (MANUAL) 

  Mattress is fully inflated and operated in manual dynamic mode with patient in direct contact with Alternating    
  cell surface (underneath top PU outer cover). Surface pressure is adjusted manually via handset to maximize    
  patient comfort (constant low-pressure therapy – CLP). This is particularly beneficial to hyper sensitive patients  
  where the benefits of the reactive therapy are matched to comfort. 

Summary of Contents for UTS

Page 1: ...USER MANUAL Universal Therapy System UTS ...

Page 2: ...reparation of Area 5 5 Installation Guide 6 Initial Set Up 6 7 6 Hand Set System Control 8 7 Maintenance Troubleshooting 9 8 Cleaning Disinfection Protocol 10 9 Specification 11 10 EMC Related Notifications 12 13 14 15 16 11 Storage Care 17 12 Waste Disposal 18 13 Symbols Used 18 14 Expected Service Life 18 15 Warranty 19 16 Legal Disclaimer 19 17 Service 20 18 Service Notes 21 19 Service Log 22 ...

Page 3: ... liquids and sharp objects 7 Never drop or insert any object into any openings 8 Do not use outdoors operate where aerosol spray products are being used or where oxygen is being administered 9 ALWAYS disconnect from power supply before opening up the mattress 10 The product has no user serviceable parts except for fuse replacement 11 If pain irritation numbness swelling or redness occurs discontin...

Page 4: ...nd shear General Safety It is important to read the information in this user manual before you use your Steiss Universal Therapy System UTS Please follow the guidelines below for your added safety and maintaining system performance Maximum patient weight is 175kg and minimum user weight is 30 kg Avoid exposing pump to liquids When cleaning do not use Phenol based substances This Steiss Universal T...

Page 5: ... remind the care team when the annual service is due 2 Hang the hand set on to the bed rail or mattress platform and plug the power cord into the mains outlet Ensure that electrical cables are safe and tidy and are not caught in the bed frame and free from obstruction The power LED will be at standby mode with amber light illuminating CAUTION it is important to routinely inspect power cable to ens...

Page 6: ... auto dynamic mode with top surface foam in position underneath Top PU outer cover 3 DYNAMIC MODE AUTO Mattress is fully inflated and operated in auto mode with patient in direct contact with alternating cell surface for maximum reactive therapy Foam top surface is stored inside the pocket beneath the dynamic cells 4 DYNAMIC MODE MANUAL Mattress is fully inflated and operated in manual dynamic mod...

Page 7: ...RNATE mode after 30 minutes of activation This Mattress features a cardio Pulmonary Resuscitation CPR valve at patient s right side at the head end In the event of cardiac arrest turn the CPR valves to the OPEN position for fast deflation of the air mattress Turn the dial to OPEN position to release air IMPORTANT please ensure that the care staff are trained and familiarized with the mattress and ...

Page 8: ...odate an individual s comfort if needed Default manual setting will be the level detected during Auto mode or at Level 2 if Auto detection is incomplete Display of LEVEL will automatically go back to MODE when not in operation for 10 seconds ALARM CODES DISPLAY When LP Low Pressure occurs system will alert user with a visual alarm displayed in LED display window and audible alarm will be triggered...

Page 9: ...l Therapy System UTS is plugged into mains Check Steiss Universal Therapy System UTS air connections are fitted securely Ensure airflow is coming out from pump Ensure connectors are securely fastened and reconnect pump air hoses if loose Ensure pump is turned on Check CPR is set to CLOSE position and is not leaking Ensure CPR valve is set to CLOSE position Replace CPR valve if an air leak is found...

Page 10: ...n 0 1 or 1000 parts per million available chlorine 9 Alternatively remove the cover and launder 95o C 203o F using normal detergents It is essential that articles be thoroughly dried after all cleaning procedures and before storage 10 Perform the following steps to clean the power unit and hose fittings Open the system and expose the pump which is housed in the corner at the foot end of the mattre...

Page 11: ...ode of operation Continuous Environment Temperature Operation 15ºC to 35ºC 59ºF to 95ºF Storage 5ºC to 60ºC 41ºF to 140ºF Environment Humididty Operation 30 to 75 non condensing Storage 30 to 90 non condensing Operation atmospheric pressure range 700 hPa to 1060 hPa Operation altitude 1017 feet to 9 843 feet 310 metres to 3000 metres Test standard IEC60601 1 IEC60601 1 2 and IEC60601 1 11 Safety s...

Page 12: ...minimum distance between portable and mobile RF communications equipment transmitters and the Steiss Universal Therapy System UTS as recommended below according to the maximum output power of the communications equipment Rated Maximum output power of transmitter W Separation distance according to frequency of transmitter metres 150 kHZ to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz...

Page 13: ...Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class B The Steiss Universal Therapy System UTS is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic Em...

Page 14: ...ation b 217 Hz 0 2 0 3 9 9 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18 Hz 2 0 3 28 28 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217 Hz 2 0 3 28 28 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 2 0 3 28 28 5100 5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 9 NOTE If...

Page 15: ... in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the separation dist...

Page 16: ...e s 0 5kV 1kV 2kV line s to earth 0 5kV 1kV line s to line s Not applicable Mains power quality should be that of a typical home healthcare and professional healthcare environment Voltage Dips short in terruptions and voltage variations on power supply input lines IEC 61000 4 11 Voltage dips 0 UT 0 5 cycle 0 UT 1 cycle 70 UT 25 30 cycles Voltage interruptions 0 UT 250 300 cycle Voltage dips 0 UT 0...

Page 17: ...rted to another location Temperature limitations 5 C 41 F 60 C 140 F Relative Humidity 30 to 90 3 Carefully place the hand set and power connection lead onto the mattress in readiness for safe storage within rolled mattress 4 Disconnect the CPR valve to allow the mattress to deflate quickly 5 Starting at the foot end motor body end of the mattress roll the unit up and use the base mount straps to ...

Page 18: ...ling facility at the end of its product life 13 Symbols Used 14 Expected Service Life The Steiss Universal Therapy System UTS pump has an expected minimum service life of two years To maintain the condition of the pump do ensure that the pump is serviced according to the schedule recommended by Steiss Do NOT use unapproved accessories or attempt to modify disassemble or otherwise misuse the Steiss...

Page 19: ...bility and effectiveness of the pressure care systems provided Internal and external research is and will always be ongoing These Descriptive Risk Levels should not be taken as prescriptive criteria Steiss mattresses cushions should be seen as an aid to care and DO NOT replace the need for good nursing care and intervention All Steiss products must be used as part of an individualised care plan wh...

Page 20: ...od working order Internal Mattress condition check these areas All T and L connectors Check for wear strain or damage Internal assembly tubing Check for wear strain or damage Connecting tube Check for wear strain or damage Cells condition including welds Check for wear strain or damage Pump Function Functional calibration Functional calibration check of pump mattress system This is a physical test...

Page 21: ... or similar professional body We recommend you check your relevant Healthcare Body website for any further information that may need to be incorporated into your checks These may be some country specific legislation re audit procedures for this kind of equipment You may need to change and adapt this procedure to suit your particular market SERVICE NOTES 21 ...

Page 22: ...e Centre All applicable service actions checked Open service actions completed Date Genuine Steiss parts used 7 First service due Steiss Authorised Service Centre All applicable service actions checked Open service actions completed Date Genuine Steiss parts used 2 First service due Steiss Authorised Service Centre All applicable service actions checked Open service actions completed Date Genuine ...

Page 23: ......

Page 24: ...24 Contents Supplier Contact Details ...