10. EMC - Related Notification
13
13
Manufacturer’s declaration – electromagnetic emissions
The
Steiss™ Universal Therapy System (UTS)
is intended for use in the electromagnetic environment (for home
healthcare and professional healthcare) specified below. The customer or the user of the
Steiss™ Universal Therapy
System (UTS)
should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The
Steiss™ Universal Therapy System (UTS)
uses RF energy only for
its internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The
Steiss™ Universal Therapy System (UTS)
is suitable for use in
all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations
flicker emissions
IEC 61000-3-3
Compliance
During test, the wired control box signal display flashes but it could return to normal status.
The device’s essential performance isn’t affected, therefore there is no concern relating to
basic safety. Please refer to the Risk Management Report for details.