Troubleshooting  

NYMCS-603

 

 
Some  problems  will  happen  in the  using  of the  microscope,  you  could  solve  them  according  the  following list. 
 

PROBLEMS

 

REASON FOR PROBLEMS

 

SOLUTION

 

I. Optical Part:

 

 

 

1.  Illumination  is  on, 
but the field of view is 
dark. 

The poor contact exists in the lamp house and 
the illumination system. 

Ensure the contact pin and the lamp holder pin work well 

The lamp bulb spoils 

change a new bulb 

The brightness adjustment knob is set too 
dark 

Adjust the knob in a proper position 

No use the appointed lamp bulb 

use the specified halogen Lamp 

6V20W

 

 

2. The edge of the 
field of view has 
shadow or the 
brightness is 
asymmetry 

The nosepiece is not in the located position 

Adjust it into the located position 

The surface of the lamp become black 

Change a new lamp bulb 

 

The filament shadow not clean up 

Move  in condenser  adjust  knob  front  and back to change 
the focus position to clean up the filament shadow. 

The surface of the lens is moldy or has 
contaminant 

 

Clean the lens 

3. Find dust and stain in 
the field of view 

There are stains on the specimen 

Change the specimen 

There are stains on the eyepiece 

Clean the eyepiece 

 
 
 
 

 

4. The image is 
defocus, 
low-resolution 

The objective damage 

Mend and correct the objective (send to 
factory for overhauling) 

The lens of the objective and eyepiece is 
moldy or have contaminant 

 

Do cleaning 

 

The opening of Aperture diaphragm and field 
diaphragm is not proper, and too much 
astigmatism. 

 

Change the opening of the aperture diaphragm 
and field diaphragm 

 

Fine focus system is broken 

Examine and repair the fine focus system(send to 
factory for overhauling) 

The objective is not in the center of the light 
path 

 

Turn the nosepiece to the located position 

 

5.  The  image  focus 
surface 

incline(one 

side  is  clear  and  the 
other side is faint) 

 

The illumination light incline serious 

Adjust the filament position ,let the light 
distributing of the field of view become symmetrical 
and bright 

The specimen don’t correctly place 

Put the specimen on the right position 

The nosepiece is not in the located position 

Turn the nosepiece in the required position 

 

6. The eyes are 
uncomfortable, the 
left and right fields 
of view is not 
coincided. 

The interpupillar distance is not correct 

Adjust the interpupillar distance correctly 

The diopter is not right 

Adjust the diopter according your sight 

 

 

Can’t adapt to binocular observation 

When look into the objective, do not stare at the specimen 
but at the whole field of view, or move the eyes away to 
see other things, then back into the objective 

 
 
 
 
 
 
 
 
 

Summary of Contents for NYMCS-603

Page 1: ...r Manual This manual is written for Metallurgical Microscope of NYMCS 603 For safety and for exerting the best performance making you familiar with the instrument entirely it is strongly recommended t...

Page 2: ...Notices 2 1 Components Name 3 2 Installation 4 2 1 Installing Diagram 4 2 2 Installing steps 5 3 Adjustment Operation 9 3 1 Adjustment System Installing Illustration 9 3 2 Operation 11 4 Technical Sp...

Page 3: ...pply voltage or it will bring a serious damage to the instrument 9 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to dis...

Page 4: ...3 STEINDORFF 1 Name of Components NYMCS 603...

Page 5: ...ing order of the spare parts The number expresses the installing process Before the installing make sure there is no dust and stain on the spare parts No marks by outside force on the surface of the s...

Page 6: ...er to change magnification in operation Clean the objective frequently At first use the 10X objective to looking for image then change another one Turn the objective till hear the Kai Kai sound make s...

Page 7: ...ert the binocular viewing head in the head of the body turn to the right place then fix up it by bolts 2 2 4 Install the eyepiece Fig 7 8 Insert the eyepiece in the eyepiece tube till end Fig 8 pictur...

Page 8: ...he socket on microscope safely Make sure be connected 3 Plug the power cord into the power source Socket safely Make sure be connected Do use the supplied power cord all the time If lost or damaged se...

Page 9: ...house and nearby parts will be very hot and will cause serious burns Please turn the main switch on O off pull up power plug and make sure the bulb the lamp house and periphery are all cool Then you...

Page 10: ...9 STEINDORFF 3 Adjustment Operation NYMCS 603 3 1 Adjustment Set Diagram Fig 18 19 Fig 18...

Page 11: ...10 STEINDORFF Fig 19...

Page 12: ...just the View field Diaphragm and Aperture Diaphragm until the two Diaphragm open size is properly Then you can get clear image When in using you need to adjust simultaneously part observe until to th...

Page 13: ...he interpupillar distance until the left and right fields of view coincide completely 3 3 6 Focusing Fig 25 1 Use the 10 objective focus to avoid the objective touch with the specimen you should raise...

Page 14: ...Adjusting the Tension Adjustment Collar Fig 27 The tightness of the tension adjustment collar has adjusted before leaving factory if finding it s loosing the mechanical stage drop itself because of de...

Page 15: ...Quadruple Nosepiece Objective Infinite plan Achromatic 4 10 20 choose and buy 40 Focus System Coaxial Coarse and Fine Focusing System Sensitivity and Graduation of Fine Focus 0 002mm Stage Double laye...

Page 16: ...focus low resolution The objective damage Mend and correct the objective send to factory for overhauling The lens of the objective and eyepiece is moldy or have contaminant Do cleaning The opening of...

Page 17: ...connect them exactly the installation of the bulb is wrong Install the bulb correctly The bulb burn out Change a new bulb 2 The bulb burn out in a high frequency Not use the specified lamp Use the re...

Reviews: