ANTI-MOTION
At the cost of measuring time, the Anti-Motion function intelligently settles the
determination of weight when the scale is in motion. The weaker Anti-Motion is,
the faster the scale displays the weight, but the longer it takes to be stable.
Anti-Motion can be set to: 0 (off), 1 (weakest), 2 (weak), 3 (normal), 4 (strong),
5 (strongest).
Press
or
to enter the Anti-motion setting.
- shows
Press
or
and
to change the value.
DYNAMIC WEIGHING
In some special application where the scale´s accuracy is not so important as the
scale´s stability for the weight reading, the dynamic weighing can be a useful
function to set the weight reading faster.
Press
or
to enter the Dynamic Weighing setting. Y--- shows
Press
or
and
to change the value.
GRAVITY
This value can be set from 0.000 to 9.999. Only adjust the Gravity when the scale
is used in a place where the acceleration of gravity is greatly different from the
place where it was calibrated.
Press
or
to enter the Gravity setting.
G----
shows
Press
or
, and then
to modify the digit. Press
or , and
to move the digit. Input the Gravity value. Press
to input the decimal point.
USER UNIT
The User Unit can be set from 0.000 to 9.999.
The User Unit is a named unit which is usually used in the user’s region, but
which is not included in the scale by default. It is a ratio to the System Unit. For
example, if the User Unit is set to 1.234 and if the System Unit is kg, then after
switching to User Unit, the scale measures the weight (1000kg), and displays the
calculated value (1234usr).
This means you have to watch up the maximal load hanging on the scale while
using the user unit, to not overload the scale because of a different displayed
value!
Press
or
to enter the User Unit setting.
U
---- shows
Press
or
, and then
to modify the digit. Press
or , and
to move the digit. Input the User Unit value. Press
to input the decimal point.
CALIBRATION
ENTERING CALIBRATION SETTINGS
Press twice to enter the Password mode.
The scale displays
. Press
or
, and
to modify the digital.
Press
or , and
, to move the digital. Input the Calibration password:
P8416.
Press
or
to confirm the input and to enter Calibration.
CAL
shows.
SYSTEM UNIT
The Scale displays the System Unit, e.g.,
, indicating that the system
unit is kg. The System Unit can be set to: kg and lb.
Press
or
to enter the System Unit setting. n--- shows
Press
or
, and
, to change the System Unit.
CALIBRATION GRAVITY
This value can be set from 0.000 to 9.999. The value 9.794 is set by default.
Press
or
to enter the Gravity setting.
G----
shows.
Press
or
, and then
to modify the digit. Press
or , and
to move the digit. Input the Gravity value. Press
to input the decimal point.
MAX. CAP
Do NOT attempt to set Max. Cap. greater than scale’s actual capacity. Overloading
causes severe harm to scale, and is very dangerous.
Press
or
to enter the Gravity setting.
00000
shows
Press
or
, and then
to modify the digit. Press
or , and
to move the digit. Input the max. capacity value. Press
to input the decimal
point.
ZERO DETECTION
Press
or
to enter the zero detection.
shows.
Keep the scale without load and press
or
to display the weight code
12345.
After the detection is finished, the scale automatically enter the Load1 Detection.
LOAD 1 DETECTION
The scale displays
.
Load a standard weight, and press
or
.
00000
shows
Press
or
, and then
to modify the digit. Press
or , and
to
move the digit. Input weight value. Press
to input the decimal point.
32