29
EN
DISPLAY BRIGHTNESS
Press or to enter Display Brightness. br||? shows.
Press or and to change brightness value.
Display Brightness can be set to:1(dim), 2(normal), 3(bright).
Dim LED brightness saves battery power dramatically.
IDLE MODE TIMING
Press or to enter Idle Mode Timing. idl?? shows.
Press
or
and
to change timing value.
Idle Mode Timing can be set to: 0(disabled), 5(5sec), 10(10sec), 15(15sec),
30(30sec), 60 (60sec). It is 30sec by default.
To maximize battery life, the scale automatically enters Idle Mode, when there’s
no action or the load is stable. In Idle Mode, the scale works in low-power con-
sumption status. Any key pressing or motion in load wakes up the scale from Idle
Mode.
Press
or
to exit User Setup.
SCALE CONFIGURATION
ATTENTION: The parameters in Scale Configuration are closely related
to the scale’s metrology performance. It is NOT recommended to change
any parameters unless you absolutely need to. In case you´re not sure,
please contact the customer service to check if it is necessary to modify
these parameters.
SCALE CONFIGURATION
ENTERING THE MENU
Press twice, enter Password mode.
The Scale displays
Input Scale Configuration password P0258 . To do
so, press
or
, then
, to modify the digital. Press
or , and
, to move the digit.
Press
or
, to confirm the input and enter the Scale Configuration.
Scale displays
.
DISPLAY RESOLUTION
The display resolution can be set to: The display resolution can be set to: 0.001,
0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, or 50.
Press
or
to enter the Display Resolution. E--- shows.
Press
or
and
to change the resolution value.
AUTO-ZERO RANGE
The Auto-Zero range can be set to: 0% (disabled), 2 (±2%FS), 3 (± 3%FS), 4
(±4%FS), 10 (±10%FS), 20 (±20%FS), 50 (±50%FS), 100 (±100%FS). It is set
to ±20%FS by default.
Press
or
to enter the Auto-Zero range setting.The scale displays
Press
or
and
to change the value.
MANUAL ZERO RANGE
The manual zero range can be set to: 0% (disabled), 2 (±2%FS), 3 (±3%FS), 4
(±4%FS), 10 (±10%FS), 20 (±20%FS), 50 (±50%FS), 100 (±100%FS).
Press
or
to enter the Manual-Zero range setting.
The scale displays the Manual-Zero range,
indicating 4% F.S.. Zero is
only allowed when the weight is within the Manual-Zero range.
Press
or
and
to change the value.
ZERO TRACKING RANGE
The Zero-Tracking range can be set to: 0 (disabled), 0.5 (±0.5e), 1.0 (±1.0e),
1.5 (±1.5e), 2.0 (±2.0e), 2.5 (±2.5e), 3.0 (±3.0e), 3.5 (±3.5e), 4.0 (±4.0e), 4.5
(±4.5e), 5.0 (± 5.0e). Enabling the Zero-Tracking will enhance the scale tempe-
rature and drift the performance!
Press
or
to enter the Zero tracking range setting.
The scale displays the Zero-Tracking range,
indicating 0.5e.
Press
or
and
to change the value.
ZERO RANGE
Zero Range defines the range that scale must fall into before activating the accu
-
mulation operation. When the load is removed from the scale, the weight left must
be lighter than the value set.
Press
or
to enter the Zero range setting.
shows
Press or
and
to change the value.
Press or and , to move the digit. Input the Zero Range value. Press
, to set the decimal point. Zero Range can be set from 0e to 3000e. It is set to
„5e” by deafult.
ZERO-SAVING
When Zero-saving is enabled, Auto Zero is disabled automatically. Scale calcu-
lates weight based on the last Zeroing action. It is disabled by default.
Press or , to enter Zero Saving.
shows.
Press or
and
, to change the status to „on” (enabled) or „off” (disabled).
30
EN