4
5
Hub
Display
S-Cell
Batteries
x 2, 3 or 4
Power-Out Cable
EQD Connector
Cable
1.
Detach the Display from the forearm by inserting a flathead screwdriver under the mounting
frame (
note
: the mounting frame will break, a replacement kit may be required).
2.
Remove the Display by feeding it back through the mounting frame hole and remove the full
S-Charge System from the forearm.
3.
On removal of the S-Charge System, care must be taken to ensure that the other items are not
damaged.
4. Once the S-Charge System is removed from the forearm, replace the faulty Hub with the new
one.
5.
Assemble as per the following steps:
Removal of old S-Charge Hub
Once connected, press the standby button on the S-Charge Display before refitting the rest of
the system. If the screen is blank, charge the batteries for 30 minutes and check again. If this
does not work; carry out the following steps:
•
Firstly, connect the S-Cells to the Hub (The Hub can accommodate 2, 3 or 4 S-Cell
batteries in one system).
•
Next, connect the Display with the ribbon cable to the Hub.
•
The system is now connected. Turn on and test by pressing the standby button on the
Display.
Note:
Only a replacement Hub, power-out cable and
EQD connector cable as circled above are provided
in this kit.