background image

D2272-02-05 CN5219 SEM310  WEB User Guide 

Notes 

1

 

Any span may be selected but full accuracy is only guaranteed for spans greater than the 
minimum recommended. 

2

 

Includes the effect of calibration, linearisation and repeatability. 

3

 

Custom characterisation is available pre-programmed at the factory. Contact your nearest 
Sales office. 

4

 

Consult Thermocouple reference tables for practical temperature ranges. 

5

 

For supply voltages over 30V, a minimum loop load of 500 

 is necessary. 

 

 

2.6 GENERAL

 

 
Input/Output Isolation 

 

500 VAC  

Time Constant  

 

 

(Filter Off) 0.5 secs (to 90 % of final value) 

Filter Factor Programmable 

  

Off / selectable between 1 and 32 seconds / or Adaptive 

Warm-up Time 

 

 

2 minutes to full accuracy 

 
ENVIRONMENTAL 
Ambient Operating Range 

 

(-40 to 85) °C 

Ambient Storage Temperature  

(-50 to 90) °C 

Ambient Humidity Range 

 

(10 to 95) % RH non condensing 

 
EMC 

 

 

BS EN 61326 

 
MECHANICAL 
Enclosure  

 

DIN standard terminal block size 

Material   

 

ABS 

Weight 

 

 

43 g 

Dimensions 

 

44 mm diameter x 24 mm 

Flammability 

 

UL94-V0 

 

 

2.7 SEM310X VERSION -  FOR USE IN POTENTIALLY EXPLOSIVE ATMOSPHERES

 

 

 

2.7.1 ATEX CERTIFICATE

 

 
The  SEM310X,  has  been  issued  with  a  EC-type  examination  certificate,  confirming compliance  with 
the European ATEX directive 94/9/EC for :- 
 
Intrinsic safety 

 

II 1 G EEx ia  IIC T4…T6.  

 
The equipment bears the Community Mark and subject to local codes of practice, may be installed in 
any  of  the  European  Economic  Area  (EEA)  member  countries.  The  SEM310X  housing  is  coloured 
light  blue  to  identify  the  equipment  as  suitable  for  Hazardous  area  use.    The  equipment  must  be 
installed  and  maintained  in  accordance  with  local  requirements  for  electrical  equipment  for  use  in 
potentially explosive atmospheres, eg EN60079-14 &  EN60079-17. This instruction sheet describes 
installation  which  conforms  with  BS  EN60079-14  &  BS  EN60079-17  Electrical  Installation  in 
Hazardous  Areas.  When  designing  systems  outside  the  UK,  the  local  Code  of  Practice  should  be 
consulted.  
 
 
 

Page 3 

 

2.7.1 ATEX MARKING

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
COMMON INFORMATION 
Manufacturer  

 

Status Instruments Ltd 

Type Number  

 

SEM310X 

 
CE marking 

 

 

0891 

 
 
Explosive Protection Marking   
 
(Type ia) Intrinsic Safety 
II   1  

 

 

Equipment Group and category 

 

 

Type of explosive atmosphere (Gas) 

EEx ia IIC T4..T6 

 

Intrinsic safety information 

TRL03ATEX21013X 

Certificate reference 

 

 

2.7.3 SPECIAL CONDITIONS FOR SAFE USE

 

 
As indicated by the Certificate Reference "X" suffix, special conditions apply for safe use for 
intrinsic safety applications.  They are as follows:- 
 
ZONES, GAS GROUPS, AND T RATING. 
 
When connected to an approved system the SEM310X may be installed in:- 
Zone 0 

 

explosive gas air mixture continuously present 

Zone 1 

 

explosive gas air mixture likely to occur in normal use 

Zone 2 

explosive gas air mixture not likely to occur and if it does, it will only occur for a 
short time. 

 
Be used in gas groups;- 
 
Group A   

propane 

Group B   

ethylene 

Group C   

hydrogen 

 
 
 

Page 4 

II 1 G EEx ia  IIC T4-T6.            TRL 03ATEX21013X

FACTORY MUTUAL: INTRINSICALLY SAFE APPARATUS
FOR CL1,DIV1,GP ABCD T4@85°C;T5@50°C;T6@40°C
FOR INSTALLATION AND MAINTENANCE SEE CONTROL
DRAWING SG4-3770-01

0891

TYPE:     Pt100
RANGE: 0-100°C

SEM310X HART

SER No. 000000 - 0001

R

www.status.co.uk

Summary of Contents for SEM310

Page 1: ...RATURE TRANSMITTER Status Instruments Ltd Status Business Park Tewkesbury Glos GL20 8FD Tel 44 0 1684 296818 Fax 44 0 1684 293746 Email sales status co uk Web www status co uk INDEX SECTION CONTENTS PAGE NO 1 0 DESCRIPTION 1 2 0 SPECIFICATION 1 5 3 0 INSTALLATION 6 7 ...

Page 2: ...L 100 to 600 25 TC Type N 180 to 1300 50 TC Type X 3 User defined Linearisation BS EN60584 01 BS 4937 IEC 584 1 multi segment Polynomials Basic Measurement Accuracy 2 0 04 FRI 0 04 RDG or 0 5 C whichever is greater Thermal Drift Span 0 01 C Cold Junction Error 0 5 C Cold Junction Tracking 0 05 C C Cold Junction Range 40 to 85 C Page1 2 3 MILLIVOLT INPUT Input Voltage Source Range 10 to 75 mV Chara...

Page 3: ...to identify the equipment as suitable for Hazardous area use The equipment must be installed and maintained in accordance with local requirements for electrical equipment for use in potentially explosive atmospheres eg EN60079 14 EN60079 17 This instruction sheet describes installation which conforms with BS EN60079 14 BS EN60079 17 Electrical Installation in Hazardous Areas When designing systems...

Page 4: ...r body direct to the input screw terminals The screw terminals have been designed to allow all connection wires to enter from an inner or an outer direction Care must be taken when locating the transmitter to ensure the ambient temperature remains within the specified operating range Fig 3 All Dimensions in mm 3 2 ELECTRICAL Connections to the transmitter are made to the screw terminals provided o...

Page 5: ...ators Temperature Probes Trip Amplifiers Signal Conditioners And many other products For further information on all products Status Instruments Ltd Status Business Park Tewkesbury Glos GL20 8FD Tel 44 0 1684 296818 Fax 44 0 1684 293746 Email sales status co uk Web www status co uk Page 8 2 1 1 Pt100 4 5 6 HART Temp Tx 4 to 20 mA T C 2 3 Power Supply Load 3 4 5 6 Pt100 3 3 3 4 6 4 4 10 to 40 VDC mV...

Reviews: