Specifications

Traverse length:

 0.2" (5mm)

Measurement length:

 0.16" (4mm) 

Traverse speed:

 0.08in/s (2mm/s)

Gauge range:

 8000µin (200µm)

Gauge resolution:

 2µin (0.05µm)

Display resolution:

 Ra. 0.4µin (0.01µm) 

Display resolution:

 1µin (0.1µm) for

Rz, Rt, Rp, Rv

Cut off:

 0.03" (0.8mm) ±15%

Uncertainty:

 4µin (0.1µm) or 5% of result, to 95% 

confidence level

Results range:

 Ra. 1600µin (40µm)

(See ISO 4288-1996)

Results range:

 Rz, Rt, Rp, Rv. 8000µin (199.9µm)

(See ISO 4288-1996)

Stylus: 

Diamond. Radius: 123µin (5µm) 

nominal, 90

o

 tip angle. Skid radius: .394" (10mm) 

nominal.

Filter type:

 2CR, (200:1)

Sample spacing:

 .025µin (1um)

Stylus Force:

 1gf (10mN) max at mid range 

Batteries: 

  Control Unit:

 (1) 3V Lithium 2450

  

Traverse Unit:

 (3) 3V Lithium 2450

Units:

 µin or µm

Operating conditions:

 50-95

o

F (10-35

o

C) 80% RH 

non-condensing

Size:

 4.92 x 3.15 x 1.5" (125 x 80 x 38mm)

Weight:

 7oz (200g)

Error Codes

E1:

 Infra Red communication error. Repeat 

operation.

E2:

 Measurement attempted when not calibrated. 

Calibrate the Traverse Unit. 

E3:

 Traverse Unit - Oscillator fault. Retry. If fault is 

persistent, contact Starrett

E4:

 Traverse Unit - Motor Positioning fault. Retry. If 

fault is persistent, contact Starrett

E5:

 Traverse Unit - NVRAM Error. Retry. If fault is 

persistent, contact Starrett

E6:

 Traverse Unit - Profile Over-range. Retry 

measurement.

E7:

 Traverse Unit - Parameter Over-range. Retry 

measurement.

E8: 

Traverse Unit - Motor Speed Setup error. Retry 

Set up procedure.

E9:

 Traverse Unit - Gain Setup error. Retry Set up 

procedure.

E10:

 Control Unit - Oscillator fault. Retry. If fault is 

persistent, contact Starrett

E11:

 Traverse Unit - Calibration information 

download error. Try again.

E12: 

Traverse Unit - Not set up. Run the Set up 

procedure.

E13:

 <Reserved>

E14: 

Traverse Unit - Electronic Measurement er-

ror. Retry. If fault is persistent, contact Starrett

Form 964    04/11    2.5C/T

TM

Digital Measurement Metrology, Inc

www.dmm.ca

1-866-960-9400

PRECISION IS OUR VISION

Digital Measurement Metrology, Inc

Summary of Contents for SR100

Page 1: ...tatus Display Parameter Status Parameter Selector Mode Selector ArDA Windows Stylus Start Buttons ON Button Control Unit Traverse Unit www dmm ca 1 866 960 9400 PRECISION IS OUR VISION Digital Measurement Metrology Inc TM Digital Measurement Metrology Inc ...

Page 2: ...ges Inserting the Batteries Remove battery covers as shown Then insert one 3V Lithium 2450 in to the Control Unit and three 3V Lithium 2450s in to the Traverse Unit as shown Then replace the battery covers After the batteries have been changed there will be a short delay when the SR100 is turned on Turning On the SR100 When the Control Unit is not used for five minutes or more it enters power save...

Page 3: ...s All parameters not just the selected param eter are transmitted to the Control Unit at the same time and temporarily stored The stored results may then be stepped through using the select buttons Low Battery Indicators Set Up Procedure A new SR100 is fully set up out of the box However if the pick up is changed or another alteration is made field setup may be required using the following procedu...

Page 4: ...munication error Repeat operation E2 Measurement attempted when not calibrated Calibrate the Traverse Unit E3 Traverse Unit Oscillator fault Retry If fault is persistent contact Starrett E4 Traverse Unit Motor Positioning fault Retry If fault is persistent contact Starrett E5 Traverse Unit NVRAM Error Retry If fault is persistent contact Starrett E6 Traverse Unit Profile Over range Retry measureme...

Reviews: