Starrett FORM 966 User Manual Download Page 12

12      FORM 966

ENGLISH

DISPLAYING MIN/MAX/TIR VALUES**

(TIR is "Total Indicator Reading")
Pressing the MIN/MAX/TIR button displays the MIN/
MAX/TIR values most recently measured. 

Press once for MIN (above right), twice for MAX 
(middle right) and three times for TIR (lower right).

Press a fourth time to exit. The MIN/MAX/TIR values 
may be zeroed by pressing the SHIFT/SET button.

LOCK MODE

*

The Lock Mode prevents intended or unintended operator modifications to 
the settings. 

Before activating the Lock function, preset other functions to the desired 
settings: Plus/Minus for spindle direction, ABS mode, PRESET Reading, 
Units, Limits setting, and MIN/MAX/TIR settings. These values are stored in 
memory. Then, lock these settings as follows:

1.  Press the SHIFT/SET then press and hold the LOCK button. 
2.  The Lock icon will turn on and stay on (see right).

Once the lock mode has been activated, the  

operator will not be able to change ABS/Normal  

Mode, Limits, Units, PRESET, or MAX/MIN/TIR, but 

will have access to the zero function and be able to view MIN/MAX/TIR readings.              

3.  To deactivate the lock mode repeat the process used to enter the mode. Press 

the SHIFT/SET then the LOCK buttons. The lock icon will turn off and stay off.

MIN/MAX/TIR

Button

*Available with Standard and Advanced Features

Press

SHIFT / SET

BUTTON

Then

LOCK

Button

to activate the Lock function

Press

SHIFT / SET

BUTTON

Then

LOCK

Button

to deactivate the Lock function

*Available with Standard and Advanced Features

Summary of Contents for FORM 966

Page 1: ...Trust Is In The Name User Manual FORM 966 ...

Page 2: ...TO LEA ATENTAMENTE ESTE MANUAL LIRE CE MANUEL AVANT D UTILISER L INSTRUMENT LEIA ATENTAMENTE ESTE MANUAL ANTES DE UTILIZAR O INSTRUMENTO 使用仪器前请阅读本操作手册 DIESES HANDBUCH VOR DER VERWENDUNG DES MESSGERÄTS LESEN LEGGERE ATTENTAMENTE QUESTO MANUALE PRIMA DI UTILIZZARE QUESTO STRUMENTO 01 18 Q ...

Page 3: ...2900 ELECTRONIC INDICATOR USER MANUAL ...

Page 4: ...eature Level and Resolution 9 Resolution Selection 9 Setting the Preset Function 10 Setting MIN MAX Limits Go No Go 10 Displaying MIN MAX TIR Values 12 Lock Mode 12 Data Output Connection 13 Battery Installation 13 Appendix A Technical Specifications 14 Appendix B Serial Output Connector 14 Available with Standard and Advanced Features Available only with Advanced Features ...

Page 5: ...RONIC INDICATOR THIS IS A STARRETT USER GUIDE FOR THE 2900 ELECTRONIC INDICATOR ALL SPECIFICATIONS IN THIS DOCUMENT ARE CORRECT AT TIME OF PRODUCTION AND ARE SUBJECT TO CHANGE PLEASE CONTACT STARRETT FOR FURTHER INFORMATION ...

Page 6: ...Do not apply lubricant to the spindle or use solvents 9 Do not disassemble or modify the indicator OPERATING INSTRUCTIONS 1 Install batteries included with indicator 2 Lightly clean the contact point 3 Secure the indicator to an appropriate holding device 4 Press the ON OFF button to turn the indicator on 5 Select the unit of measure inch or millimeter by pressing the IN mm button Not applicable t...

Page 7: ...ENGLISH Stem Cap Data Cable Spindle Bellows Spindle Contact Point Stem Data Output Cover Battery Drawer Back Lug 1 21 32 42 2mm 1 4 6 5mm 1 4 6 5mm 2 5 64 53mm 3 8 9 5mm INDICATOR COMPONENTS DIMENSIONS AND LAYOUT ...

Page 8: ...the display 6 LOCK Prevents operator modification to the Settings Press SHIFT SET button then press and hold the LOCK button after setting the other functions 7 RES Resolution First select the Unit English Metric Press SHIFT SET then the RES Plus Minus buttons Press the RES button to scroll through choices and press the SHIFT SET to make the selection 8 LIMITS Sets the Min and Max values for the G...

Page 9: ...0 2 09981 Basic English Metric 0001 001mm 5 12mm 2900 4 09983 Standard English Metric with Selectable Resolution 0005 0001 00005 01mm 001mm 5 12mm 2900 6 09985 Advanced English Metric with Selectable Resolution 0005 0001 00005 01mm 001mm 5 12mm 2900 1M 09986 Basic Metric 001mm 12mm 2900 4M 09988 Standard Metric with Selectable Resolution 01mm 001mm 12mm 2900 6M 09990 Advanced Metric with Selectabl...

Page 10: ...T button and repeat step 3 Repeat steps 2 and 3 until all the digits are SET 5 Tosetanegativevalue pressPRESETwhentheplus minussigniconisflashing 6 To exit the PRESET function press the PRESET button when the PRESET icon is flashing The SET value will remain displayed 7 Press and hold the ZERO ABS button to return to the active measuring mode SETTINGMIN MAXLIMITS GO NOGOFUNCTION 1 Select the units...

Page 11: ...eyond the set limits the display will flash and the MIN or MAX LIMIT icon will indicate the direction the limit have been exceeded To exit Press the LIMITS button Press SHIFT SET BUTTON Then LIMITS Button to enter Min Limit Mode Press SHIFT SET BUTTON The LIMIT icon will flash on off Press SHIFT SET BUTTON to increment the value of the digit or the plus minus sign Then press SHIFT SET BUTTON To ca...

Page 12: ... MIN MAX TIR settings These values are stored in memory Then lock these settings as follows 1 Press the SHIFT SET then press and hold the LOCK button 2 The Lock icon will turn on and stay on see right Once the lock mode has been activated the operator will not be able to change ABS Normal Mode Limits Units PRESET or MAX MIN TIR but will have access to the zero function and be able to view MIN MAX ...

Page 13: ...e plug as shown below When not using output replace the data cover to protect the electronics BATTERY REPLACEMENT Remove the two Phillips screws that secure the battery drawer as shown below Remove the battery drawer and discard the old batteries Place two new CR2032 batteries into the tray with the positive side facing up Slide the tray carefully into the indicator and secure with the two screws ...

Page 14: ... displacement stored MX Displacement is minimum peak displacement stored TR Displacement is the T I R stored LL Displacement is the lower or minimum limit setting LU Displacement is the upper or maximum limit setting HL Displacement is a held displacement 1 Sign character minus but no plus 1 Space character IN mode only 6 Digits 1 Decimal point 2 Spaces 1 Character strings to indicate units IN or ...

Page 15: ...l Order Only State Change High to Low SERIAL INPUT VOLTAGE LEVEL SPECIFICATION Logic 0 Vin 1 30 Volts Logic 1 Vin 1 98 Volts SERIAL OUTPUT VOLTAGE LEVEL SPECIFICATION MINIMUM MAXIMUM Low Level Vss Vss 0 25 1 5mA Load High Level Vdd 0 25 1 25 mA Load Vdd Note Vss 0 Volts Vdd 2 7 to 3 6 Volts ENVIRONMENTAL CONSIDERATION Temperature 10 30 C Humidity 30 85 RH no condensation Atmosphere Non corrosive N...

Page 16: ...Starrett com ...

Reviews: