Example: Configuration Number
Example: Firmware Version
Calibrate the Tool
Page 4
4.0 New battery, Startup Sequence
•
Each time you change the battery(Pg.7), the micrometer will go through a startup sequence. At the end of the
sequence you will need to move the spindle to initialize the measurement system. When the micrometer
comes on the display will go through a series of changes as shown in Fig.1-5.
Fig.3
Fig.1
Fig.2
Full Lamp Display
Fig.4
•
When “CAL” is displayed, Move the thimble very slowly in one direction, at a rate of .25 (1/4th) of an inch of
movement over the span of 2 -3 seconds, until the display changes to show measurement readings. This
movement will initialize the accuracy tuning function and calibrate the indicator.
•
This feature will happen each time the batteries are changed.
•
If you don’t move the spindle, “CAL” will stay on the display.
If you move the spindle to fast it will take longer
for the tool to initialize.
5.0 Zero Sequence
•
After “CAL” mode is completed press the
ZERO/ABS
button and the display will show a dashed line as
seen in Fig.6A below. The line will incrementally disappear from left to right Fig.6B; this is a visual clue
to show how long you should wait for the tool to zero out. Make sure not to move the spindle during that
time. This will happen each time the tool is zeroed, and takes less than a second to complete.
Fig.6A
Fig.6B