page 25 of 36
Prior to autoclaving set the hand controller
to “autoclaving”.
Prior to autoclaving turn cap (1) and remove
(2).
Cleaning and decontaminating
Autoclaving
The hand controller with adapters and aspira-
tion hose is designated for steam sterilization at
121°C and 2 bar absolute (1 bar overpressure).
Time of exposure according to DIN 58946 t
e
=
20 minutes.
DO NOT exceed a temperature of 121°C.
Comply with instructions for use of the manu-
facturer of the autoclave.
It is the user´s responsibility to ensure effective
autoclaving.
NOTICE
NOTICE
It is not possible to disassemble the hand con-
troller completely.
It is possible to disassemble and replace the
adapters, the screw cap and the aspiration
hose, see “Accessories and spare parts”.
If the housing opens slightly by mistake, e. g., if
the handle is dropped, press the gap together
by hand. Perform a function test.
NOTICE
1
2
121°C
2 bar absolute