17
77-496 / 77-429 / 77-439 / 77-497 / 77-427
exit Calibration Mode.
• Axis settings are now saved, Calibration Mode is OFF,
and laser tool begins Auto-Levelling as done when
initially powered ON.
NOTE:
• A press of
will slope the axis by 3,5 arc
seconds. Reference Figure
L
for resulting slope direction
for each key.
• If the laser tool can still not be calibrated after
following the Calibration procedure, please send
the laser tool into an Authorized Service Center for
repair.
Vertical Check
(See Figure
P
)
(Only necessary on models with Vertical Auto-Levelling)
• Set the laser tool on a stable surface in its vertical
position 1 m away from a wall that extends ≥ 2 m high
with the ”Y1” side facing that wall (
P
1
).
• Power ON the laser tool and allow the laser tool to
Auto-Level and be sure laser is rotating.
• Mark reference points “A”
(where laser line is on
fl oor 1 m away from wall)
, “B “
(where laser beam
is at corner)
, and “D
1
”
(where laser beam is 2 m
up the wall)
.
• Rotate the laser tool 180° so that the “Y2” side is now
facing the wall (
P
2
).
• Align the laser beam with reference points “A” and “B”
and then go back to the wall and measure the distance
between the reference point “D
1
” and “D
2
” (
P
3
).
• There is no need to adjust calibration if the distance
between reference point “D
1
” and “D
2
” is < 1,0 mm.
• If the distance measured is ≥ 1,0 mm then a calibration
adjustment is necessary.
Vertical Calibration
(See Figure
P
)
(For RL HW+ substitute
where ever
is
referenced below)
• With laser tool powered OFF, press and hold
followed by
.
• Release
and continue to hold
for ≥ 3
seconds.
• Release
.
• The LED/LCD will indicate laser tool is in Calibration
mode.
• If necessary, adjust the vertical “X” axis by pressing
until the laser beam is aligned with “D
0
”.
“D
0
” is the halfway point between
“D
1
” and “D
2
”
during vertical axis check.
• .
• Press
to set the vertical “X” axis.
• Axis setting is now saved, Calibration Mode is OFF,
and laser tool begins Auto-Levelling as done when
initially powered ON.
NOTE:
• A press of
will slope the axis by 3,5 arc
seconds. Reference Figure
L
for resulting slope direction
for each key.
• If the laser tool can still not be calibrated after
following the Calibration procedure, please send
the laser tool into an Authorized Service Center for
repair.