page 12 of 12
26.
After all 10 range / coupling combinations are calibrated, power cycle the
instrument. When the lock-in turns on, with the 50 Ω or shorting cap still
installed and the frequency at 3.0 Hz (set by the Offset Cal function), verify the
displayed “R” value is less than 1.0 mV on all settings of Input Range and
Coupling.