ST&G 1320A Instructions Manual Download Page 7

 

 

CE Conformity 

This  product  meets  the  requirements  of  93/42/EEC  guidelines  for  medical  products. 

This  product  has  been  classified  as a  class  I  product  according  to  the classification 

criteria outlined in appendix IX of the guidelines. Please keep this manual in safe place 
for future use. 

7.1 

Have the patient pre-scheduled for a 2-week follow-up appointment, as the brake sensitivity and 

the knee movement may change due to initial wear in period. It is recommended that readjustment 
of the knee joint’s settings again be verified after the patient has spent a period of time getting used 
to the prosthesis after another 3-4 weeks. 

 

7.2 

It is recommended to perform routine annual safety checks. Should issues arise in regards to 

parts or repairs, please contact with your distributor or ST&G USA Corp. 

 

7.  Service and Repair 

ST&G USA Corporation Warranty applies only when the product is used according to the specified 

conditions and for the intended purposes, following all manufacturer’s recommendations. We offer a 
warranty of 2 years 

Summary of Contents for 1320A

Page 1: ...Instructions for 1320A Pneumatic Single Axis Knee with Weight Activated Lock ST G USA Corp Phone 714 524 0663 2691 Saturn St Fax 714 364 8113 Brea CA 92821 www stngco com i 01 ...

Page 2: ...2 1 Figures Fig 1 ...

Page 3: ...3 Fig 3 Diagram for reference only may not reflect exact knee Fig 2 ...

Page 4: ...is 9 and with the Lower Joint Section 12 through the Knee Axis 2 The slotted Swing Block 4 accepts the slotted Brake Bushing 3 which surrounds the knee axis During weight bearing pressure on the Upper Joint Section 6 and the Plastic Block 7 transfers the load on the Swing Block 4 which increases the resistance between the knee axis and brake bushing increasing the stability of the knee In the unlo...

Page 5: ...iction and brake effect are cohesively adjusted Take care in making individual adjustments and reassess function in the parallel bars Important Do not use talcum powder Talcum powder reduces the lubrication of the mechanical parts which may lead to a malfunction and thus increase the risk of failure Using this medical product after application of talcum powder will render all claims against ST G n...

Page 6: ...w 3mm hex Knee Flexion Damping Valve Adj screw 3mm hex Important Be aware that the brake sensitive adjustments and pneumatic valve adjustments will compliment each other and can affect the patients walking speed This can occur by the pneumatics speed force input affecting the latest brake sensitivity adjustment performed so it is important to verify through visual and verbal patient feedback durin...

Page 7: ...e due to initial wear in period It is recommended that readjustment of the knee joint s settings again be verified after the patient has spent a period of time getting used to the prosthesis after another 3 4 weeks 7 2 It is recommended to perform routine annual safety checks Should issues arise in regards to parts or repairs please contact with your distributor or ST G USA Corp 7 Service and Repa...

Page 8: ...8 ST G USA Corporation www stngco com e mail info stngco com 2691 Saturn Street Brea CA 92821 USA Tel 1 714 524 0663 Fax 1 714 364 8113 1320A IFU Rev B 10 18 17 ...

Reviews: