15
Incorrect settings
…
…
can be dangerous to life, depending on the
application. For this reason, the following
instructions are of particular importance:
Before you use the wrench, check that the needle
is on zero. If the wrench is not under load and the
needle is not exactly over the zero mark, refer to
the section headed "What to do if the needle is not
at zero" on page 20.
Additional safety instructions
…
…
will depend on the actual situation. You will find
these in the sections below marked with the
following symbol:
ã
.
Care of the tool
Manoskop
®
No 71 is a precision tool and must be
treated as such. Avoid subjecting the tool to jolting
and keep away from chemicals and sources of
heat over and above the intended purpose.
Please note that extremes of temperature and
humidity can cause the readings displayed to be
falsified.
Avoid subjecting the tool to loads exceeding 20 %
above the rated maximum either in the direction of
tightening or in the opposite direction. This may
damage the Manoskop
®
No 71.
Max. load:
value + 20 %