Table 11.
Guidance and manufacturer’s declaration –
electromagnetic immunity
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Recommended
separation
distance
d=0.58
√
P
where P is the
maximum output
power rating of
the transmitter in
watts (W)
according to the
transmitter
manufacturer
and d is the
recommended
separation
distance in
meters (m).
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
80% AM at 1
kHz
20 V/m
Recommended
separation
distance
d=0.18
√
P 80 to
800 MHz
d=0.35
√
P 800
MHz to 2.5 GHz