background image

 

© 2010 SST Sensing Ltd                                        2                                             www.sstsensing.com 
DS0074 Rev 7 

DESCRIPTION 

The O2I-FLEX Interface provides all necessary          
electronics to power and control SST Sensing’s range of 
dynamic Oxygen Sensors.  
 
The O2I-FLEX Interface can be user configured to output 
measuring ranges of 0-25% O

and 0-100% O

. The    

entire measurement range is linear in both cases. Factory    
default is 0-25% O

2. 

When configured for 0-100% O

the 

user can also customise the analogue output ranges to 
suit their application. 
 
The interface outputs the measured values                  
simultaneously via 4 output channels, 4-20mA, 0-10V

DC

 

and RS232 (Rx and Tx), all outputs are referenced to the 
system GND.  
 
A digital 3.3V

DC 

logic output cycles at the same  

frequency as the electrochemical pumping action of the 
oxygen sensing cell during normal operation, thus  
providing a real time sensor health check, if the output 
ceases to cycle the sensor has entered a start-up or error 
state. This provides fault proof operation. The digital  
output is also used during the calibration process to  
indicate the interface status. 
 
A green on-board LED mirrors the CYCLE output and can 
be used to visually determine the sensor status or during 
the calibration process. A red LED indicates the unit has 
power applied. 
 
SST’s range of oxygen sensors do not directly measure 
the oxygen concentration but instead measure the partial 
pressure of oxygen within the measurement gas. In order 
to output an oxygen concentration (%) the O2I-FLEX 
must be calibrated, or more specifically, re-referenced in 
a known gas concentration, typically normal air. 
 
Calibration, or re-referencing, is achieved by connecting 
the calibration input to GND or by pressing the on-board 
calibration switch and monitoring the status of the digital 
cycle output or by visually monitoring the on-board green 
LED. During the calibration process the output will either 
automatically calibrate to a fixed reference or can be 
manually calibrated to any output by way of a PCB 
mounted potentiometer. The fixed reference is factory set 
to 20.7% O

2

 for calibration in normal air though this value 

may be altered via the RS232 interface for calibration 
with a reference gas of any known oxygen concentration. 
Calibration is stored on power loss.  
 

 

The auto or manual calibrate function is user  
configurable. Regular calibration removes the effects of 
application and atmospheric pressure changes and also 
eliminates any sensor drift that may occur during the first 
few hundred hours of operation.

 

 
For  more  detailed  information  on  the  operation 
of  SST  Sensing  Oxygen  Sensors  please  refer  to 
the following application note via our website:  
 
AN0043 Operation Principle and Construction of 
Zirconium Dioxide Oxygen Sensor. 

 
 
 
 

O2I

-

Flex

 Oxygen Interface 

Summary of Contents for O2I-Flex

Page 1: ...lectable output filtering allows adaptive fast and dynamic or slow and stable output response Power and Sensor Operating LEDs Removable polarised screw terminals for easy wiring SPECIFICATIONS Maximum ratings Supply voltage 24VDC 10 Current consumption 600mA max 24VDC 4 20mA Load 100 600Ω Temperature limits Storage 10 to 70 C Operating 10 to 70 C O2I Flex APPLICATIONS Combustion control including ...

Page 2: ...ygen sensors do not directly measure the oxygen concentration but instead measure the partial pressure of oxygen within the measurement gas In order to output an oxygen concentration the O2I FLEX must be calibrated or more specifically re referenced in a known gas concentration typically normal air Calibration or re referencing is achieved by connecting the calibration input to GND or by pressing ...

Page 3: ...ial pressure of oxygen PPO2 within the measurement gas deviations in the Barometric Pressure BP from that present during calibration will cause readout errors proportional to the change For example if the sensor was reading 21 O2 at 1013 25mbar and the BP increased by 1 the sensor readout would also increase by 1 to 21 21 O2 OXYGEN SENSOR HEATER ADJUSTMENT The Oxygen Sensor heater must be adjusted...

Page 4: ...0 SST Sensing Ltd 4 www sstsensing com DS0074 Rev 7 OUTLINE DRAWING AND MOUNTING INFORMATION 60mm 60mm 67mm 67mm 4 x M4 Mounting holes on industry standard 60x60mm pitch PCB LAYOUT O2I Flex Oxygen Interface ...

Page 5: ...ery SST oxygen sensor has two heater connections which should be connected to pins 1 2 of the O2I Flex the heater coil has no polarity However when connecting to a sensor where the sensor housing is one of the heater connections pin 1 of the O2I Flex should be connected to the housing SYSYEM BLOCK DIAGRAM PIN Description 1 Sensor Heater GND Yellow H 2 Sensor Heater Yellow H 3 Sensor Sense Blue S 4...

Page 6: ...onnections ensure Tx goes to Rx of the PC and Rx goes to Tx of the PC The O2I Flex communicates via standard COM port settings that are default on most PCs and many other RS232 compatible devices If however communication problems are occurring use the settings below to configure the PC or device COM Port O2I Flex Oxygen Interface ...

Page 7: ...ith the amount of averaging user variable whilst the raw un averaged values allow the user to detect sudden oxygen changes The averaged value is the measurement output on both the 4 20mA and 0 10VDC outputs The sensor Td value is the measure of the partial pressure of oxygen in the measurement gas The O2 concentration is the Td value scaled by the stored calibration value To stop or restart the da...

Page 8: ...onse to oxygen changes Press Enter to save The new averaging value is now stored in memory This value is retained on power loss Adjusting the Minimum and Maximum Ranges of the Analogue Outputs 4 20mA and 0 10VDC The O2I FLEX is factory default to output a range of 0 25 O2 via its two analogue outputs This range can be expanded to 0 100 O2 as described on Page 6 When the unit is reconfigured to out...

Page 9: ...umed At this point remove GND from PIN 10 or release the calibration switch The output will now track to the correct value for the calibration gas Calibration is complete Calibration values are retained on power loss Manual Calibration Ensure the O2I FLEX is configured for manual calibration See CONFIGURATION on Page 6 Place the sensor probe in the calibration gas typically normal air Allow the ou...

Page 10: ...ng from Water Droplets In environments where falling water droplets are likely the sensor should be protected from water falling directly onto the very hot sensor cap as this can cause massive temperature shocks to the cell and heater Popular methods include a hood over the sensor cap or for the sensor to be mounted in a larger diameter cylinder At a very minimum the sensor cap should be angled do...

Page 11: ...te O2 measurement is required Investigated gases were H2 Hydrogen up to 2 stoichiometric combustion CO Carbon Monoxide up to 2 stoichiometric combustion CH4 Methane up to 2 5 stoichiometric combustion NH3 Ammonia up to 1500 ppm stoichiometric combustion 2 Heavy Metals Vapours of metals like Zn Zinc Cd Cadmium Pb Lead Bi Bismuth will have an effect on the catalytic properties of the Pt electrodes E...

Reviews: