SSD Statiron DZ4 Instruction Manual Download Page 8

6.    Point the ion balance plate at the 

measurement location. 

 

 

The potential on the ion balance is 

measured.   

 

 

NOTE 

  If  the  display  panel  blinks  "1",  the  charged  voltage  has  exceeded  the 

measurable range (over-range). If this occurs, stop measuring immediately, as 
this may cause damage to the meter. 

 

7.    The value on the display panel is the measurement result (unit: 

kV).   

 

8.    In I.B mode, you can temporarily hold the measured value by 

pressing the hold button. 

 

If you press the hold button again, the stored value is removed. 

You cannot use the hold function in MAX I.B mode. 

 
9.    To turn the meter off, press the power button for more than 1 

second.   

 

 

 

Summary of Contents for Statiron DZ4

Page 1: ...lectrical potential meter with digital readout for measuring the electrical potential of charged objects When measuring fluctuating static electricity you can temporarily hold an indicated value and c...

Page 2: ...osed to water or oil Maintenance Regularly remove any built up dirt etc from the optional ion balance plate Built up dirt can cause insulation faults Make sure to turn the main power of the device OFF...

Page 3: ...using it for the first time Instruction manual warranty x 1 this document Main Device Earth Lead x 1 Strap x 1 Battery 006P type 9V alkaline dry cell battery x 1 Soft case Model number ODZ4 SFTCS x 1...

Page 4: ...al changes according to the field strength concentrated in the electrode of the sensor portion of the meter and the size of the charged object In particular if the charged object is smaller than the t...

Page 5: ...ad to a grounded object To obtain accurate measurements the earth terminal must be grounded Measurement Procedure Measuring Static Electricity 1 Turn the meter on Press the POWER button for more than...

Page 6: ...re 3 Fig 3 Red LED mark NOTE If the recommended measuring distance of 30mm cannot be attained do not bring the meter any closer if the potential displayed exceeds 19 99kV Doing so may cause damage to...

Page 7: ...e the I B mode 3 Zero point adjust the meter Point the sensor at a grounded object and press the 0ADJ button NOTE Zero point adjustment is reset when you turn the meter off 4 Attach the ion balance pl...

Page 8: ...s occurs stop measuring immediately as this may cause damage to the meter 7 The value on the display panel is the measurement result unit kV 8 In I B mode you can temporarily hold the measured value b...

Page 9: ...ed on the upper right of the display panel When the battery charge icon shows one unit left replace the battery Error Display The signal detection method of the meter is an oscillating chopper method...

Page 10: ...correct Remedy 1 Perform zero point adjustment again Cause 2 Structural components close to the sensor are charged Remedy 2 Wait until the charge in the components has attenuated The display does not...

Page 11: ...measured and sensor Measuring distance adjustment Red LED beam focusing method Sensor head rotation angle 180 stops every 45 Display LCD display with built in backlight Polarity display polarity pola...

Page 12: ...where you purchased the product SHISHIDO ELECTROSTATIC LTD http www shishido esd co jp HEAD OFFICE MARUNOUCHI BLDG 9F 918 4 1 MARUNOUCHI 2 CHOME CHIYODA KU TOKYO 100 6309 SALES DEPARTMENT TOKYO BRANCH...

Reviews: