SSCOR DUET 2314 Series Operating Instructions & Maintenance Manual Download Page 7

CHARACTERISTICS

 

SPECIFICATIONS

 

Size

 

17"L x 9"H x 5.25"W

 

(43.18cm L x 22.86cm H x 13.33cm W)

 

Weight

 

10.65 lbs (device only)

 

4.83 kgs (device only)

 

Vacuum Pump

 

12V DC.

 

Clinical Airflow 

≥ 

30LPM

 

Exceeds 525mmHg

 

Variable Regulator (7)

 

Controls negative pressure

 

Regulator Gauge (8)

 

Calibrated in mmHg. Color

-

coded

 

 

Power Source: 

 

             Battery (DC Power) 

 

             Medical Switcher (AC Power)

 

 

Rechargeable Sealed Lead Acid.  SSCOR part # 80635

 

100V

-

240VAC, 47

-

63Hz Uses 1ea. 3A 250V fast acting fuse

 

Switch (5) 

 

Off/Standby/On membrane switch

 

AC receptacle (1A)

 

Uses (2) 2.5A 250V fast acting fuses

 

Collection Canister (13)

 

1200cc/ml   SSCOR part #48041

 

Patient tubing

 

Vinyl tubing 

9/32”

ID x 

72”

L

 

(7.1mm ID x 182.88 cm L)

 

SSCOR part #43200

 

Suction Tip

 

HI

-

D® 

Big Stick®

” 

SSCOR part #44241C

 

7

 

Clearing The Airway Is Our #1 Priority

 

General Specifications

 

SSCOR DUET 

-

 Model 2314 Series

 

Rev Y (11/18) 

 

Summary of Contents for DUET 2314 Series

Page 1: ...g The Airway Is Our 1 Priority Operating Instructions Maintenance Manual Tel 1 818 504 4054 Tel 1 800 434 5211 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com shop sscor com Email sales sscor com techsupport sscor com ...

Page 2: ...tandard e g IEC 60950 for IT equipment and the IEC 60601 series for medical electrical equipment In addition all such combinations systems shall comply with the standard IEC 60601 1 1 safety requirements for medical electrical systems Equipment not complying with IEC 60601 shall be kept outside the patient environment as defined in the standard Any person who connects external equipment to signal ...

Page 3: ...SSCOR Duet is a portable 100V 240V AC constant suction device with a 12V DC battery back up The battery is charged by an internal DC dual mode battery charger The charger monitors the battery charges the battery only when necessary shuts down the unit if the battery is low and signals battery condition A fully charged battery at full capacity will power the unit for 45 minutes 10 SSCOR warrants th...

Page 4: ...ping Carton Manufacturer Shipping Carton Date of Manufacture Shipping Carton Authorized Representative in the European Community Control Panel Power Source Connected Control Panel Pump is On Control Panel Battery Level Of Charge Green LED Full Green Yellow LED High Yellow LED Half Yellow Red LED Low Red LED Battery Depleted Put device back on charge Control Panel Push On Push Off 4 Clearing The Ai...

Page 5: ...gauge located above the vacuum regulator The vacuum reading should rise to 300mmHg from zero in 3 seconds It should max out at approximately 525mmHg Lower levels of negative pressure will be observed at altitude 3 Adjust the vacuum reading to your desired level If the vacuum does not meet or exceed 525mmHg check for a leak in the system i e tubing barb connections or loose canister lid Refer inter...

Page 6: ... desired setting turn the device off and put it back on charge SSCOR recommends replacement of the battery after 3 years Replacement battery may be purchased at shop sscor com PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Power Source Connected LED not lit Fuse blown Molex connections disconnected Power cord disconnected Replace fuse in power supply or input receptacle Reconnect Molex Reconnect cord Do...

Page 7: ...ery DC Power Medical Switcher AC Power Rechargeable Sealed Lead Acid SSCOR part 80635 100V 240VAC 47 63Hz Uses 1ea 3A 250V fast acting fuse Switch 5 Off Standby On membrane switch AC receptacle 1A Uses 2 2 5A 250V fast acting fuses Collection Canister 13 1200cc ml SSCOR part 48041 Patient tubing Vinyl tubing 9 32 ID x 72 L 7 1mm ID x 182 88 cm L SSCOR part 43200 Suction Tip HI D Big Stick SSCOR pa...

Page 8: ...e main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system E PC Board Electrical Circuits Internal Dual Mode Battery Charger Do not attempt to service the PC Board Return to SSCOR or authorized service center for service F Power Supply Do not attempt to service the Power Supply Return to SSCOR or authorized service center for service G Functional gro...

Page 9: ...t filter is soiled remove the two screws holding the filter door in place see page 8 remove the soiled filter and replace with new filter SSCOR part number 90160 Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Disinfection Use personal protective equipment such as gloves a smock and face and eye...

Page 10: ...will be secure at a force equal to or greater than 10G Vertical Mount Remove the four 20 x 1 1 2 hex head screws from hardware packet Mark the hole placement for a verti cal mount and use a 5 16 drill to drill the holes Coat the screws with Loctite not supplied before securing the Charging Retention Bracket Put the four 1 1 2 screws through the holes in the charging bracket Model 8323 Reattach the...

Page 11: ...t the end of the run test turn the device off and reconnect the device to the charging source For a further explanation of the SSCOR battery run test visit the SSCOR Website at http www sscor com training_videos html 3 Determine if the gauge is properly calibrated After the battery has been fully charged when the battery condition LED on the control panel is green connect tubing from the patient p...

Page 12: ... D If the tubing from the device to the canister is clamped to the device 4 remove the tubing from the patient port of the canister and take the reading at the distal end of the tubing 5 The airflow at the distal end of the tubing should be at least 48 LPM E If the device still does not meet the airflow specification call SSCOR technical support 1 800 434 5211 or 1 818 504 4054 to trouble shoot Af...