SSCOR DCell Suction – Model DM10-001
Rev. D (02/17)
7
Clearing The Airway Is Our #1 Priority
Trouble Shooting
Problem
Possible Cause
Corrective Action
Does not function when
switch is on (DC Power)
Battery discharge
Replace Batteries
Molex connections disconnected Lift battery cover and check to see
Molex connections are secure
Defective PC Board
Replace PC Board
Fuse Blown
Replace Fuse
Red LED
Low battery capacity
Replace Batteries
Battery condition indicator
lights not lit
Defective PC Board
Replace PC Board
No suction when pump is
running
Vacuum line is disconnected
Re-connect vacuum line
Canister Defective
Replace canister
Canister lid loose
Tighten canister lid
Catheter thumb vent is open
Occlude thumb vent
Defective pump(s)
Replace pump(s)