SSCOR DCell Suction Operating Instructions & Maintenance Manual Download Page 6

SSCOR DCell Suction – Model DM10-001 

Rev. D (02/17) 

Clearing The Airway Is Our #1 Priority 

Operating Instructions (Continued) 

Remove the discharged batteries from the 12xD Cell battery holder and dispose of the batteries ac-
cording to local / regional / national requirements for the disposal of alkaline batteries. Install 12 new fresh 
D Cell alkaline batteries into the 12xD Cell battery holder - being careful to assemble the batteries in the 
correct orientation as indicated on the battery holder. When the battery holder is loaded with 12 new D 
Cell alkaline batteries, set the 12xD Cell battery holder back into the chassis. The battery holder only fits 
in one orientation. Re-connect the Molex on the battery holder to the wiring harness. Turn the device on 
to verify the batteries are correctly installed and the battery holder is connected to the wiring harness. Ob-
serve the battery voltage indicator on the control panel. The green LED should be lit indicating the batter-
ies have a high level of charge.  
 

Running the device using the #80665 DC power cord (Available for purchase on shop.sscor.com)

 

Connect the DC power cord to the 12V DC cigarette lighter receptacle in the vehicle. Connect the DC 
power cord to the suction unit by securely attaching the charging plug into the receptacle.      If the suction 
unit is operated while it is hooked up to the vehicle it will utilize the vehicle power and save its own battery 
for emergency use. The DC power cord will not charge the 12xD Cell batteries when connected to the 
suction device.  
 
CAUTION: The #80529 power adapter will not charge the 12xD Cell batteries when connected to the suc-
tion device. If the 80529 power adapter is connected to the SSCOR DCell Suction device and the battery 
voltage indicator shows a red LED, the output from the power adapter is not sufficient to power the 
SSCOR DCell Suction device. The device is running from its internal batteries. The 80529 power adapter 
may be purchased at shop.sscor.com. 

Do not store batteries in the SSCOR DCell Suction device if it is not expected to be used for 
several months. 

WARNING:  

To avoid risk of shock, this equipment must only be connected to a supply mains with  
protective earth.  
 
The 80529 Power Adapter is for Indoor Use Only.  
 
WARNING:  
Periodic checking or replacement of “D” Cell Alkaline batteries is required.  
 
If for any reason poor battery quality is suspected, perform a battery test as described below.  

BATTERY TEST: 

 

Turn the SSCOR DCell Suction device on and run the unit for 30 seconds. Occlude the patient port 
of the canister and observe the negative pressure gauge. The negative pressure gauge should rise 
above 300mmHg in 3 seconds and reach 525mmHg. Observe the battery condition indicator 
LED  

    

.  If it lights red, it is time to install 12 fresh D Cell batteries into the SSCOR DCell Suction 

battery holder. Dispose of the batteries according to local /regional / national requirements for the 
disposal of electronic waste.  

Summary of Contents for DCell Suction

Page 1: ...nance Manual The world s 1st and only full size portable suction device powered by ALKALINE BATTERIES Tel 1 818 504 4054 Tel 1 800 434 5211 Fax 1 818 504 6032 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com shop sscor com Email marketing sscor com techsupport sscor com ...

Page 2: ...ot store batteries in the SSCOR DCell Suc tion device if it is not expected to be used for sev eral months 9 Keep battery contact surfaces and battery com partment contacts clean by rubbing them with a rough cloth each time you replace batteries 10 Make sure you insert batteries into your device properly with the plus and minus terminals aligned correctly Always install batteries per the labeling ...

Page 3: ...ons on battery replacement For replacement parts and accessories please visit shop sscor com The SSCOR DCell Suction device is designed for hospital crash carts patient transport and emergency medical service Suction power can be regulated when full power may be considered harmful to the pa tient SSCOR suction units are designed to provide instant effective suctioning independent of exter nal sour...

Page 4: ... Label Battery Voltage Indicator Serialized Label Separate Collection for Electronic Equipment Serialized Label Type BF Equipment Serialized Label Attention Consult Accompanying Documents Shipping Carton Authorized Representative in the European Community Shipping Carton Date of Manufacture Shipping Carton Manufacturer Shipping Carton Transport Storage Range Shipping Carton Humidity Shipping Carto...

Page 5: ... should be replaced with fresh D Cell Alkaline batteries To access the battery holder open the top of the chassis exposing the battery holder 1 Disconnect the Molex on the 12xD Cell battery holder from the wiring harness 2 Lift the 12xD Cell battery holder out of the chassis noting the orientation of the battery holder The battery holder only fits in this orientation Do not lift the battery holder...

Page 6: ...utilize the vehicle power and save its own battery for emergency use The DC power cord will not charge the 12xD Cell batteries when connected to the suction device CAUTION The 80529 power adapter will not charge the 12xD Cell batteries when connected to the suc tion device If the 80529 power adapter is connected to the SSCOR DCell Suction device and the battery voltage indicator shows a red LED th...

Page 7: ...ver and check to see Molex connections are secure Defective PC Board Replace PC Board Fuse Blown Replace Fuse Red LED Low battery capacity Replace Batteries Battery condition indicator lights not lit Defective PC Board Replace PC Board No suction when pump is running Vacuum line is disconnected Re connect vacuum line Canister Defective Replace canister Canister lid loose Tighten canister lid Cathe...

Page 8: ...40 00001 Wiring Harness Uses 1 2A 250V slow acting fuse UL Listed and CSA Certified Type 1 x Collection Canister 1200cc ml SSCOR Part 48041 Patient Tubing Vinyl tubing 9 32 ID 72 L SSCOR Part 43200 Suction Tip HI D Big Stick Large Bore Suction Tip with thumb control SSCOR Part 44241 Operating Temperature Range 0 C 32 F 40 C 104 F Operating Relative Humidity 0 93 non condensing Operating Atmospheri...

Page 9: ...ean using a mild detergent and if necessary disinfect with a mild disinfectant such as 10 to 1 bleach and rinse using clear water to remove any residue Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Disinfection Use personal protective equipment such as gloves a smock and face and eye protectio...

Page 10: ...eter of the back chassis Next remove the two screws holding the D Ring of the carry handle on each side of the device The device will then open To close the device replace the screws and tighten Make sure all pneumatic and wiring connec tions are secure before closing the chassis A 1 2 3 4 5 1 2 B C D A PC BOARD ELECTRICAL CIRCUITS Do not attempt to service the PC Board Return to SSCOR for service...