Model DM10-001 

Clearing The Airway Is Our #1 Priority 

Operating Instructions 

& Maintenance Manual 

The world’s 1st and only full size portable suction  

device powered by ALKALINE BATTERIES. 

Tel 1+818-504-4054 
Tel 1+800-434-5211 
Fax 1+818-504-6032   

               , 

INC.

 

11064 Randall Street 
Sun Valley, CA 91352  USA 
 
www.sscor.com 
shop.sscor.com 
Email: [email protected] 
[email protected] 

Summary of Contents for DCell Suction

Page 1: ...nance Manual The world s 1st and only full size portable suction device powered by ALKALINE BATTERIES Tel 1 818 504 4054 Tel 1 800 434 5211 Fax 1 818 504 6032 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com shop sscor com Email marketing sscor com techsupport sscor com ...

Page 2: ...ot store batteries in the SSCOR DCell Suc tion device if it is not expected to be used for sev eral months 9 Keep battery contact surfaces and battery com partment contacts clean by rubbing them with a rough cloth each time you replace batteries 10 Make sure you insert batteries into your device properly with the plus and minus terminals aligned correctly Always install batteries per the labeling ...

Page 3: ...ons on battery replacement For replacement parts and accessories please visit shop sscor com The SSCOR DCell Suction device is designed for hospital crash carts patient transport and emergency medical service Suction power can be regulated when full power may be considered harmful to the pa tient SSCOR suction units are designed to provide instant effective suctioning independent of exter nal sour...

Page 4: ... Label Battery Voltage Indicator Serialized Label Separate Collection for Electronic Equipment Serialized Label Type BF Equipment Serialized Label Attention Consult Accompanying Documents Shipping Carton Authorized Representative in the European Community Shipping Carton Date of Manufacture Shipping Carton Manufacturer Shipping Carton Transport Storage Range Shipping Carton Humidity Shipping Carto...

Page 5: ... should be replaced with fresh D Cell Alkaline batteries To access the battery holder open the top of the chassis exposing the battery holder 1 Disconnect the Molex on the 12xD Cell battery holder from the wiring harness 2 Lift the 12xD Cell battery holder out of the chassis noting the orientation of the battery holder The battery holder only fits in this orientation Do not lift the battery holder...

Page 6: ...utilize the vehicle power and save its own battery for emergency use The DC power cord will not charge the 12xD Cell batteries when connected to the suction device CAUTION The 80529 power adapter will not charge the 12xD Cell batteries when connected to the suc tion device If the 80529 power adapter is connected to the SSCOR DCell Suction device and the battery voltage indicator shows a red LED th...

Page 7: ...ver and check to see Molex connections are secure Defective PC Board Replace PC Board Fuse Blown Replace Fuse Red LED Low battery capacity Replace Batteries Battery condition indicator lights not lit Defective PC Board Replace PC Board No suction when pump is running Vacuum line is disconnected Re connect vacuum line Canister Defective Replace canister Canister lid loose Tighten canister lid Cathe...

Page 8: ...40 00001 Wiring Harness Uses 1 2A 250V slow acting fuse UL Listed and CSA Certified Type 1 x Collection Canister 1200cc ml SSCOR Part 48041 Patient Tubing Vinyl tubing 9 32 ID 72 L SSCOR Part 43200 Suction Tip HI D Big Stick Large Bore Suction Tip with thumb control SSCOR Part 44241 Operating Temperature Range 0 C 32 F 40 C 104 F Operating Relative Humidity 0 93 non condensing Operating Atmospheri...

Page 9: ...ean using a mild detergent and if necessary disinfect with a mild disinfectant such as 10 to 1 bleach and rinse using clear water to remove any residue Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Disinfection Use personal protective equipment such as gloves a smock and face and eye protectio...

Page 10: ...eter of the back chassis Next remove the two screws holding the D Ring of the carry handle on each side of the device The device will then open To close the device replace the screws and tighten Make sure all pneumatic and wiring connec tions are secure before closing the chassis A 1 2 3 4 5 1 2 B C D A PC BOARD ELECTRICAL CIRCUITS Do not attempt to service the PC Board Return to SSCOR for service...